
22
Using the RespSense II Monitor
3. Replace the single-use, disposable moisture trap and filter before each use. Refer to
Single-Patient Use, Disposable Moisture Trap and Filter
for instructions on how to replace
the moisture trap and filter.
4. Connect the sample line to the connector on the moisture trap and secure it by turning the
Luer lock connector clockwise. Only use sample lines recommended by Nonin (see
).
5. See
Monitoring a Patient
1. Verify the system has been set up (see
2. Attach the sample line to the patient, as described in
, or refer to
the individual sample line instructions for use.
3. Press
On/Standby
to turn the monitor on.
4. Monitor runs a self-test (see
).
5. Select alarm limits. If alarm limits are locked, this screen does not display.
6. Verify the graphs and alarm limits display on the touch screen display.
7. If allowed, adjust the alarm limits for the patient. Refer to
for
instructions on how to change alarm limits.
8. The audible alarm function activates approximately 2 minutes after start-up. The operator
can activate the audible alarm before the 2 minutes are up by pressing the Audio Pause
button.
9. The monitor is ready to use. The patient can stay connected to the monitor for as long as
needed.
CAUTION:
After exposing the monitor to an environment outside of normal room
temperature/humidity conditions, always replace the moisture trap and filter before
each use.
WARNING: Prior to connecting the monitor to the power supply and a power outlet, be
sure to verify the voltage and frequency rating on the power supply are the same as
the outlet. If this is not the case, do not connect the monitor and power supply to the
outlet.
WARNING: The monitor displays a flashing yellow battery indicator (low battery) when
it has approximately 60 minutes of use remaining before it shuts itself off.
WARNING: When turning on the monitor, verify that a beep is heard. If a beep is not
heard, do not use the device. The speaker may not be functioning properly.
CAUTION:
Each time the system is turned on, audible alarms are silenced for
2 minutes unless the operator presses the Audio Pause button.
CAUTION:
The patient’s nasal passage may dry out if continuous monitoring is
required. Check patient hourly for nasal comfort.
!
!
!