18
Display Screens
3
RR BrPM
Displays the respiration rate in breaths per minute.
4
EtCO
2
Messages
Example:
Shows capnometer alarm messages. See
for more
information.
5
Audible Alarm
No symbol means audible alarms are enabled.
A bell with dashed lines indicates that audible alarms are
paused.
A bell with solid lines indicates that audible alarms are off.
6
Respiration Graph
Displays a graph of the CO
2
in exhaled air (capnograph).
7
Full Partial
Low Critical Charging
Battery
The battery indicator shows the approximate percentage of
battery life remaining.
•
Full/Partial
– Battery indicator is green.
•
Low
– Battery indicator flashes yellow.
•
Critical
– Battery indicator flashes red.
•
Charging
– Battery indicator displays a lightning bolt
when the monitor is connected to the power supply
and a power outlet.
NOTE:
When the monitor reaches a low or critical battery
condition, an audible alarm sounds if the audible alarm is on.
To clear the alarm, connect the monitor to the power supply
and a power outlet.
8
EtCO
2
Trend Graph
Scale is 20 – 80 mmHg (2.7 – 10.7 kPa). This scale is fixed
and cannot be changed.
9
RR Trend Graph
Scale is 5 – 35 BrPM. This scale is fixed and cannot be
changed.
Table 4. Operating Screen – Display Descriptions (Continued)
No.
Symbol/Name
Description