25
The upper alarm limit is always located above the displayed value, and the lower limit is always
located below the displayed value.
NOTE:
The monitor will always reset the alarm limits to the factory default settings once it is
turned off and turned on again.
Alarms
The intended operator’s position for correctly perceiving a visual alarm signal and its priority is
1 meter (3.3 feet).
Alarm Function
An alarm activates under certain conditions, such as if an alarm limit is outside the set limits, if
the patient is not connected, or if an equipment fault occurs. The alarm is both visual (a blinking
parameter, limit, or a message) and audible (beeping tones at different intervals).
If an alarm activates when the trend screen is displayed, the monitor returns to the operating
screen. When an alarm is active, the trend screen cannot be accessed.
Alarm Silence
There are three categories of LifeSense alarms: pulse oximetry, capnography, and system.
Audible alarms can be temporarily disabled using the Audible Alarm Pause/Resume button .
Decreases upper limit
→
Decreases lower limit
→
←
Increases upper limit
←
Increases lower limit
CAUTION:
Set or adjust only one parameter at a time.
Pulse Oximetry Alarms
Capnography Alarms
System Alarms
SpO
2
(high and low)
Pulse rate (high and low)
NO PROBE
NO FINGER
ARTIFACT
SIGNAL LOW
EtCO
2
(high and low)
Respiration rate (high and low)
NO BREATH
OCCLUSION
TRAP FULL? / PUSH ALARM
BATT LOW
DISP ERROR
NO OXIMET
NO CAPNO
!