Introduction 

The NONIN 8500AV Veterinary Pulse Oximeter continuously monitors arterial hemoglobin oxygen saturation and 
pulse rate.  Its intended application is monitoring animals. 

The Model 8500AV Oximeter is intended for spot-checking and displaying numerical values for arterial oxyhemoglobin 
saturation and pulse rate. 

The Model 8500AV Oximeter is small (3.0"W x 6"H x 1"D) and lightweight (10 oz. – 280 gm).  It will typically operate 
for 100 hours continuously between battery replacements. 

The Model 8500AV determines arterial oxyhemoglobin saturation (%SpO

2

) by measuring the absorption of red and 

infrared light passed through the tissue. Changes in light absorption caused by pulsation of blood in the vascular bed are 
used to determine arterial saturation and pulse rate.  The Model 8500AV requires no routine calibration or maintenance. 

Oxygen saturation and heart rate are displayed on light emitting diode (LED) digital displays. On each detected pulse, 
the perfusion LED flashes. Patient perfusion signals are graded as good, marginal, or inadequate and are indicated as 
such by the LED changing color from green to yellow or red. This simple method gives the user a pulse by pulse visual 
indication of waveform signal quality without requiring the user to perform complex waveform analysis during critical 
patient care situations.  

Sensor disconnect or malfunction is indicated by lack of good perfusion flashes and/or a minus (-) sign to the left of the 
SpO

2

 value on the LED display.  Ultimately, if good perfusion pulses are not received, the SpO

2

 and Pulse Rate 

numerical will be replaced by dashes. When the batteries are low, the digital displays will blink. 

Содержание 8500AV

Страница 1: ...Operator s Manual Model 8500AV Alarm Hand Held Veterinary Pulse Oximeter Nonin Medical Inc 13700 1st Avenue North Plymouth MN 55441 5443 763 553 9968 1 800 356 8874...

Страница 2: ...recautions for Use 1 Introduction 3 Operating Instructions 4 General 4 Connect Sensors 4 Turn on the Oximeter 4 Verify Operation 4 On Alarm Models Only 4 Veterinary Sensors 5 Sensor Placement 6 Batter...

Страница 3: ...carefully to verify normal operation The use of accessories sensors and cables other than those listed in this manual may result in increased emission and or decreased immunity of this device Caution...

Страница 4: ...all equipment must be installed and put into service according to the EMC information specified in this manual This device has not been tested for immunity to electromagnetic disturbances Portable and...

Страница 5: ...determine arterial saturation and pulse rate The Model 8500AV requires no routine calibration or maintenance Oxygen saturation and heart rate are displayed on light emitting diode LED digital display...

Страница 6: ...the perfusion light If the perfusion light is red or yellow reposition the sensor or try a different sensor If normal operation cannot be achieved call NONIN Customer Support for assistance After eigh...

Страница 7: ...nitoring These sensors are manufactured to meet the calibration requirements for the NONIN Model 8500AV Pulse Oximeter The oximeter is calibrated for adult human hemoglobin measured at the finger tip...

Страница 8: ...easurement error Secure the sensor with tape ensuring that the tape does not restrict perfusion Model 2000T Transflectance Sensor Application The 2000T Transflectance sensor is designed for placement...

Страница 9: ...the left hand display indicating the Alarm mode dft will appear in the right hand display Each time the arrow switch is pressed the option on the right hand display will advance When the correct valu...

Страница 10: ...h nn 00 through 12 Day d 01 through 31 Hours h 00 through 23 Minutes nn 00 through 59 Printer mode is used by the model 8500P Printer WARNING The use of the 8500P may result in increased emission and...

Страница 11: ...hat may affect the SpO2 data The following is a simple illustration of how to interpret the perfusion LED Perfusion LED blinks GREEN pulse waveform signal is of good quality and SpO2 data is accurate...

Страница 12: ...splayed or when the unit is out of track Press the 1 switch while holding the arrow switch This will cause the current settings of the four alarms to be displayed in the same format and order as durin...

Страница 13: ...Cathode 3 Infrared Cathode Red Anode 4 Serial Data TTL Levels 5 Detector Anode 6 Logic Level 7 Cable Shield 8 Coax Shield 9 Detector Cathode 5V a Serial format Real time mode ASCII 9600 BAUD 9 Data B...

Страница 14: ...cale for adults using or finger clip sensors Below 70 not specified b Rate Accuracy 3 1 digit 7 Temperature a Operating 20 to 50 C b Non Operating 30 to 50 C 8 Humidity a Operating 10 to 90 Non Conden...

Страница 15: ...e work should be done under a microscope set at 25 times magnification Use of regular soldering equipment will damage the board and void any applicable warranty Nonin will repair out of warranty oxime...

Страница 16: ...systems or accessories delivered to the purchaser which are found to be defective in any manner whether such remedies be in contract tort or by law This warranty excludes cost of delivery to and from...

Страница 17: ...er Model 8500MB Mounting Bracket Wall or Pole mount Model 8500CC Carrying Case Black or Yellow Model 8500I Cable 3 foot extension Model 8500VI Veterinary Cable 10 foot extension Model 8500PCC Printer...

Отзывы: