16
Operator Functions
The Model 7500 features a number of advanced options, which are
intentionally more difficult to activate. These functions are recommended
only for trained operators and they require multiple button presses to
prevent accidental activation.
Function
Button
Instruction
Recall Previous
Alarm Limit
Settings
Press the Limits button while the unit is on.
“
rCL
” appears, indicating that previous
alarm limit settings may be recalled. To
recall the settings, press the Plus button
and select “
yES
.” Press the Limits button
again to confirm.
Memory
Playback
Press and hold the Plus (+) button while
turning on the Model 7500. This functions
with the NONIN nVISION® software.
Select the Model 2500 option in nVISION
software (the Model 2500 option also
applies to the Model 7500FO).
NOTE
: Alarm limits cannot be changed when the Model 7500 is in Patient
Security mode. Patient Security mode prevents accidental changes to critical
parameters. The Model 7500 allows users to lock and unlock alarm limits,
volume settings, and time settings
Enter Patient
Security Mode
To enter Patient Security mode, press and
hold the Alarm Silence button while turning
on the device.
Exit Patient
Security Mode
To exit Patient Security mode, press and
hold the Alarm Silence and Limits buttons
while turning on the device.
+
+
+
+