11
Operating Modes and Defaults
Operating Modes and Defaults
The Model 7500 features Setup mode, Factory Defaults, User-Defined
Defaults and Patient Security modes.
NOTE:
Patient Security mode overrides any default settings.
Setup Mode, Viewing Limits and Setting Time
In Setup mode, users can adjust alarm limits and volumes, set clock and
calendar information and clear the device’s memory. Pressing the Limits
button activates Setup mode, and all adjustments can be made using the
Plus (+) or Minus (-) buttons. Setup mode is available when the device is
operating, or during the startup/initialization process. Time is set by
setting each of the last five options in setup mode: year, month, day, hour
and minute.
Setup mode is not available in Patient Security mode. In Patient Security
mode, pressing the Limits button scrolls through the limits on the displays,
allowing the operator to view the current limits. Pressing and holding the
Plus (+) button also reviews the limits, regardless of operating mode.
Factory Defaults
In Factory Defaults, all adjustable parameters are set as indicated in the
table below. This is the Model 7500’s default operating setting.
The Model 7500 is shipped with factory defaults active. To revert to
Factory Default alarm limits from the User-Defined Default alarm limits,
simultaneously press the alarm silence and minus (-) buttons.
NOTE:
User-Defined Default values are lost when Factory Defaults are set
active.
Default alarm and volume settings are automatically selected for every
operating session in which the parameters were not recalled or changed
within the setup menu.
Alarm Limit
Factory
Default
Adjustment
Options
Increment
SpO
2
High Alarm Limit
Off
Off, 80-100
1%
SpO
2
Low Alarm Limit
85%
Off, 50-95
1%
Pulse Rate High Alarm Limit
200 BPM
Off, 75-275
5 BPM
Pulse Rate Low Alarm Limit
50 BPM
Off, 30-110
5 BPM
Alarm Volume
High
Off, Low, High
N/A