Nonin 3100 WristOx Скачать руководство пользователя страница 40

40

No pulse display on 
bargraph, cont’d.

Finger is cold.

Warm the finger.

Finger is wet.

Dry the finger and inside of sensor.

Indicator not lit in finger 
insertion area.

Call NONIN Customer Support.

Excessive patient motion.

Reduce patient motion.

Pulse Oximeter Sen-
sor indicator appears.

Poor signal detected from 
finger.

Reposition or replace finger; keep 
sensor motionless. 
Warm the application site.

The WristOx needs repair.

Call NONIN Customer Support.

Problem

Possible Cause

Possible Solution