TEMP./MEMORY key

Activates the temperature 

selection function. Key is also 

used to scroll through the saved 

measurements.

(1) Needle holder

 

Place the needle (5) in the holder when 

not in use.

(2) Main display

 

Shows the measurement result and the 

temperature setting.

(3) Connector for external power supply/

battery charger

  See “External power supply/battery 

charger” for details.

(4) Collected measurements display

  Shows the number of a stored 

measurement. The corresponding 

measurement result

 

and temperature is displayed in the main 

display (2). See “Memory function” for 

details.

(5) Needle
(6) Filter/Water trap

  Protects the device from particles and 

liquids.

(7) Pressure relief valve

 

Use to perform a pressure relief of the test 

bottle before measuring.

Содержание NomaSense CO2 P2000

Страница 1: ...User Guide ...

Страница 2: ...ice Safety instructions Personnel operating and maintaining the device must be familiar with all aspects of its operation and be proficient in maintenance Such personnel should review the following precautions to promote safety awareness Refer to the User Guide before operating or maintaining the device Do not open the device In case of technical problems please contact your service provider Do no...

Страница 3: ...perature 0 40 C Humidity 95 RH non condensing CO2 Sensor Non dispersive infrared NDIR temperature compensated Range 0 2 5 g L Resolution 0 01 g L Accuracy 50 mg L abs 0 1 g L 100 mg L abs 1 2 5 g L Expected life time 3 years Measuring system Measuring time 10 sec Sample volume approx 23 mL Built in pump max 200 mbar vacuum expected life time 500 hours 180 000 measurements Calibration 0 CO2 and 100...

Страница 4: ...cs switches the device into TEST MODE MEASURE key Starts a measurement cycle of 10 seconds Pressing the key during a measurement will extend the measurement with another 10 seconds The key is also used to select the proper measurement temperature when function is activated by the key Pressing and holding this key while pressing the TEMP MEMORY key will start a sensor calibration Overview 6 7 3 4 2...

Страница 5: ...nector for external power supply battery charger See External power supply battery charger for details 4 Collected measurements display Shows the number of a stored measurement The corresponding measurement result and temperature is displayed in the main display 2 See Memory function for details 5 Needle 6 Filter Water trap Protects the device from particles and liquids 7 Pressure relief valve Use...

Страница 6: ...e setting range is 0 25 C in steps of 5 C When required temperature setting has been selected press the ON OFF key to save the setting and return to measuring mode this will happen automatically after app 10 secs if the ON OFF key is not pressed 3 Fill 50 ml of wine in the test bottle 8 then insert the rubber septum 10 and finally tighten the lid 9 4 Shake the test bottle vigorously for app 10 sec...

Страница 7: ...memory is full Clear memory If the memory is full it must be cleared before new values can be stored Press and hold TEMP MEMORY key After 2 seconds the memory is cleared and 0 is shown in the display 4 indicating that the memory is empty Show collected measurements Press TEMP MEMORY key twice and the first collected measurement is shown The small display 4 shows the measurement number flashing and...

Страница 8: ...utlet you must fit the appropriate adapter 1 Select the appropriate adapter 11 from the kit delivered with the device 2 Fit the adapter to the power supply 12 by inserting it at the top and then pushing it in at the bottom until it clicks 3 To remove the adapter simply press the lock 13 downwards and hold it while removing the adapter Battery charging When the display 2 shows Lo BAt the battery ne...

Страница 9: ...nt If there is any dirt or fluid in the measuring gas hose 17 it can be cleaned using dry compressed air CAUTION When using compressed air for cleaning the hose must be disconnected from the device Replacing needles filters and valves During use of the device particles or liquids may over time be sucked through the measuring gas system Therefor it will be necessary to replace both the needle 5 the...

Страница 10: ... small display 4 When measuring CO2 the bottom of the display 1 shows CO2 If measured value is negative the display will show CO2 CO2 followed by a decimal point The device will be in TEST MODE until it is switched OFF Performing 0 calibration Prepare the device with a proper 0 CO2 calibration gas Press and hold the MEASURE key Now press TEMP MEMORY key then release both keys Throughout the calibr...

Страница 11: ...10 ...

Страница 12: ... Contact Us for your Customized Solution Nomacorc LLC 400Vintage Park Dr Zebulon NC 27597 U S A Tel 1 919 460 2200 nomacorc com Nomacorc ZI Les Plénesses Chemin de Xhénorie 7 B 4890 Thimister Clermont Belgium Tel 32 87 63 88 20 nomacorc com nomacorc com ...

Отзывы: