background image

 

7 - Sense LWEU Series 

 

Specifications 

Environment 

Storage conditions 

 

-40…+60°C, relative humidity 5-95%, non-condensing

 

Operating conditions 

 

Sense: -20...+45°C, relative humidity 0-100% 

 

 

Sense IR: -20...+45°C, relative humidity 5-95%, non-condensing 

Protection class 

 

IP65 

Enclosure material 

 

PC+ABS (Bayblend® T88-4N) 

 

Dimensions 

Dimensions 

 

Sense: 224 x 28 x 23 mm 
IR: 119 x 28 x 23 mm 

 

 
Temperature measurement probe (Sense) 

Probe 

 

80 mm long, 2 mm diameter, hollow stainless steel probe with sharp tip 

Sensor 

 

Pt100 sensor element inside of the probe tip 

Measuring range 

 

-50…+200°C (theoretical Pt100 linearized range -200...+600°C) 

Accuracy 

±0,5 °C within range -30…+100 °C, when device temperature is +5…+45°C and 
stable 

Measuring interval 

About 5 seconds 

 
IR temperature measurement (Sense IR) 

Sensor 

Gradient-compensated thermopile infrared temperature sensor with integrated 
optics 

Mesurement beam 

 

12°, practical measurement distance 5…20 cm 

Target emissivity 

 

Fixed 0,95 

Measuring range 

 

-70…+380°C 

Accuracy 

±1 °C within range -20…+100°C, when device temperature is +5…+45°C and 
stable 
±2 °C within range -40…+120°C, when device temperature is +5…+45°C and 
stable 

Measuring interval 

 

About 5 seconds 

 

Radio connection 

Description 

 

LoRaWAN standard, EU 868 MHz frequency range 

Transmitter module 

 

Murata ABZ-093 LoRaWAN modem 

Antenna 

 

Internal 

Compatibility 

  

Nokeval NSnappy Hubs and Cloud Service 

Frequency 

 

863-870 MHz (LoRaWAN 1.0.2 EU band) 

Power 

 

Max +14 dBm E.R.P. 

Range 

Depends on environment, indoors: in good conditions up to hundreds of meters, 
outdoors: line-of-sight up to 10 km 

Power supply 

Type 

 

3.6 V lithium thionyl chloride primary battery, not user replaceable 

Battery life 

 

At least 1 year with 20 minutes of use per day 

 
 
 
 

Содержание NSnappy Sense LWEU Series

Страница 1: ...3004891141 V4 16 3 2022 Firmware 1 0 1 0 User manual Sense LWEU Series...

Страница 2: ...h Sense IR 6 The reading only updates with a few second intervals 6 Specifications 7 Warnings 8 Trademarks 8 Manufacturer 8 Introduction Nokeval NSnappy Sense series measuring instruments are LoRaWAN...

Страница 3: ...iew 3 Either select Enter new measurement or start performing a task that contains entering a measurement result 4 Turn on the device by briefly pressing its power button 5 Blue LED light on the devic...

Страница 4: ...fore the first readings appear on the screen Measuring with NSnappy Sense IR Point the lens of the IR sensor directly towards the target object Measuring distance should be between 5 to 20 cm The smal...

Страница 5: ...sive cloth moistened with a mild detergent solution Be careful not to scratch the IR sensor s lens Do not use any abrasive devices or substances when cleaning Do not clean the device in a dishwashing...

Страница 6: ...could give incorrect readings The readings appear to be too high or too low with Sense IR 1 Make sure that you are pointing the lens directly to the target 2 Make sure that you are measuring from as c...

Страница 7: ...urement Sense IR Sensor Gradient compensated thermopile infrared temperature sensor with integrated optics Mesurement beam 12 practical measurement distance 5 20 cm Target emissivity Fixed 0 95 Measur...

Страница 8: ...ike in a parked car where temperature can rise to over 50 C The battery may overheat and cause fire or explosion hazard Read this manual carefully before using the device Only Nokeval authorized servi...

Отзывы: