Thermo Scientific AquaSensors
™
AnalogPlus
™
Dissolved Oxygen Sensor (ppm) User Guide
10
3.2.
Sensor Construction
The AnalogPlus Dissolved Oxygen sensor has 1-inch NPT threads on the front and back of the
sensor body for easy installation in tee, union mount and immersion mount hardware.
Save the protective cap and use it to
keep membrane moist whenever the
sensor is removed from service.
If the sensor does not calibrate, refer
to Section 4 to review maintenance,
refurbishment and troubleshooting
options.
The sensor should be positioned at
least 15
°
up from horizontal. This
will insure that entrained air does not
form a pocket around the sensor
head.
Dimensions: IN (mm)
DO NOT USE PIPE SEALANT.
Pipe sealants may not provide
adequate sealing or may react with
different plastic materials. Use
thread tape.
DO NOT OVER TIGHTEN!
Maximum torque: 65 lbs
⋅
inch
(7 N
⋅
m)
Figure 3.2: 1 Inch Sensor Body