Relocation Of Guide Rail
If it is necessary to relocate the guide rail and sensor assembly use the following procedure.
1. Remove complete assembly from the pipe.
2. Insert a small screwdriver in the hole at the end of the guide rail moulding and lever up the clip
holding the electronics assembly by pressing down on the screwdriver as shown below.
3. Repeat 2 on the other end and then pull off the electronics unit.
Figure 21
4. Disconnect the sensors.
5. Remove the original Gel pads from the sensors.
6. Push the sensor blocks into the guide rail to enable the washers and locking screws to be
refitted.
7. Replace the Gel pads. A spare set is supplied with the unit and further sets are available from
Micronics Ltd.
8. Follow the original procedure for installing the guide rail on the pipe.
Figure 20 Diagnostics Menu
Instrument Type:
Heating
Installation:
Flow Side
Fluid Type:
Water
Set type of energy exchange, either Heating or Chiller.
Set location of the flow measurement, either Flow or
Return.
Set fluid, either water or 30% glycol