41
4ZTUFN4QFDJĕDBUJPOT
Result Range:
1.1 - 33.3 mmol/L
Sample Size:
Minimum 0.5 microlitre (0.5 μL)
Sample:
Fresh capillary whole blood,
venous blood drawn in sodium or lithium
heparin tubes, or TRUEresult Control Solution
Test Time:
Results in as little as 4 seconds
Result Value:
Plasma values
Assay Method:
Electrochemical
Power Supply:
One 3V lithium battery
#CR2032
(non-rechargeable)
Total power when active
at full battery = 8.6mW
Battery Life:
Approximately 2146 tests or 1.5 years
Automatic shut-off:
After two minutes of non-use
Weight:
25.5 g w/out battery; 28.3 g with battery
Size:
8.4 cm x 3 cm x 15.2 cm
Memory Size:
500
blood
glucose
results,
1 Control Test result
Operating Range
(Meter & Test Strips):
Relative
Humidity:
10-90% (Non-condensing)
Temperature:
10°C - 40°C
Haematocrit:
25-60%
Note:
Use
within
specified
environmental
conditions
only.
Chemical Composition
TRUEresult Test Strips:
Glucose dehydrogenase-FAD
(Aspergillus sp.)
, mediators, buffers and stabilisers.
TRUEresult Control Solution:
Water, d-glucose,
buffers, viscosity enhancing agent, salts, dye and
preservatives.