TRUE METRIX
®
PRO System Comprehensive Resource Guide (GDH-FAD Enzyme)
Quality Control Testing 18
Quality Control Testing
• Use ONLY TRUE METRIX
®
Control Solution to
perform Control Tests.
• Write date opened on control solution
label. Discard if either 3 months after first
opening or after the date printed next to
the EXP on the bottle label has passed,
whichever comes first.
• Control Test ranges are printed on the
TRUE METRIX
®
PRO Test Strip vial label.
3 Levels of Control Solution
Quality Control Testing is used to detect errors that
may occur due to test system errors, product defects,
adverse environmental conditions and variance in
operator performance. Ongoing Quality Control Testing
is also used to detect any performance issues of the
system over time. Facility Quality Control Testing Policy
and Procedure should adhere to the manufacturer’s
instructions for use and regulatory guidelines.
TRUE METRIX
®
PRO is a no-coding system, which means
the meter does not have to be coded to each lot of test
strips. To assure accurate and reliable results, TRUE
METRIX
®
PRO offers two kinds of Quality Control Tests.
These tests ensure that the TRUE METRIX
®
PRO System
is working properly and the user’s testing technique
is good.
May 30,
201
7
Содержание True Metrix Pro
Страница 3: ...Introduction...
Страница 9: ...System Components...
Страница 16: ...Clinical Information...
Страница 21: ...Meter Setup Quality Control Testing...
Страница 29: ...Blood Glucose Testing...
Страница 32: ...Diabetes Data Management Capabilities...
Страница 37: ...System Maintenance...
Страница 43: ...Training Program...
Страница 53: ...References...
Страница 55: ...Appendix A Clinical Studies Note Available for download at www niprodiagnostics com...
Страница 57: ...Appendix C Safety Data Sheets...
Страница 60: ...Appendix D Forms...
Страница 62: ...Appendix E Proficiency Testing Information...