15.
The electronic pipette has a self-check function. When there is an error in these results, the pipette
displays a following error massage “Err XX”. When an error can not be released even after applying a
following remedy to the pipette, please apply for a repair. (Refer to “15. WHEN REQUESING REPAIRE”.)
Circumstances
Considerations for cause
Remedies
“
Err 01
”
The automatic position resetting is
not working.
If the connection nut (Refer to
“7-5. Parts names”) is loose,
retighten it.
Press the Reset key to reset the
pipette.
“
Err 02
”
Step motor error.
Press the Reset key to reset the
pipette.
The
sample
liquid
lingers
inside
the
pipette tip.
Compatibility
in
characteristics
between the sample liquid and
pipette tip.
Select the reverse operation.
Use a tip that matches the
sample liquid.
The sample liquid
leaks
from
the
pipette tip, or the
aspirated volume is
small.
A wet tip was used.
Use a new tip.
The tip is improperly attached.
Attach the tip properly.
Quality error of the pipette tip.
Use a high quality tip.
Sealing problem in the piston area.
Replace the lower part of the
pipette.
Wear on the tip holder.
Replace the tip holder.
Aspirating error
Weak battery.
Recharge the battery.
The tip ejector is improperly
attached.
Disassemble
and
then
reassemble the lower part of the
pipette.
The tip holder is clogged.
Remove the clog.
The piston movement is not
smooth.
Clean the piston.
Replace the lower part of the
pipette
The power cannot be
turned on.
Contact failure of the battery.
Set the battery again.
Dirt on the battery terminal.
Replace the battery with a new
one.
Battery discharge.
Recharge the battery.
Deterioration of the battery.
Replace the battery with a new
one.