EU Patient Manual
–560-IFU-IT1098, Rev. A
Firmware
Error
The system fails
to operate in a
controlled fashion.
Note:
The
infusion will be
stopped
automatically
.
The screen
displays
“Firmware Error”;
the LED flashes a
red light; the audio
alert sounds.
Press any key to
acknowledge the alert and
power off the pump.
Take the pump out of
service. Use a replacement
pump if needed.
Battery
Depleted
The infusion will
continue for 30
minutes before
the device powers
off.
The screen
displays “Battery
Depleted” for 5
seconds and
repeats displaying
every 20 seconds;
the empty battery
level indicator
symbol flashes;
the LED shows a
solid yellow light;
the audio alert
sounds.
Press the RUN/STOP key
to suspend the infusion.
Press the ON/OFF key for 3
seconds to power off the
pump.
Replace the battery
following the instructions
provided in Section 4:
Battery Replacement in this
user manual.
Power Off and
Replace
Battery
This battery has
been used for
over 240 hours
/infusion volume
has reached 1500
mL. The pump
has reached the
end of its
performance life.
The screen
displays “Power
Off and Replace
Battery”, the LED
flashes a red light;
the audio alert
sounds.
Press the RUN/STOP key
to acknowledge the
condition and silence the
audio alert.
Replace the battery
following the instructions
provided in Section 4:
Battery Replacement in this
user manual