Getting
started
44
XTM0011-F2
XT V 160 X-ray System
5.5.1
The Inspect-X computed tomography workflow
To make the acquisition of a Computed Tomography (CT) volume easy, the settings and
parameters that are required are logically organised on tabs that define the CT workflow in
Inspect-X.
To begin, click
CT
to select the CT workflow.
Start
This is the starting point of CT acquisition and allows you to choose the method of
acquisition. The available methods are:
•
New Circular CT Scan
Keeps the parameters in the workflow at their current values. Select this option to run a
circular CT scan on a new sample or sample type. You then need to define the
parameters by working through the tabs in the CT workflow.
•
Use Parameters of Last Scan
This option is useful if you want to repeat an acquisition using the same sample or a
sample of the same type run under the same conditions.
•
Use Archived CT Profile
Uses parameters from an archived CT profile. CT profiles can be archived; if multiple
versions of a profile are available, you can choose which version you want to use.
•
Select CT Profile From Dataset
Sets the CT processing parameters to those in a selected CT profile dataset. When you
load a CT profile, the manipulator is moved to the required position, the X-ray conditions
are set up, and the imaging device parameters are reloaded.
Position and Optimise
This tab is used to position the sample for CT and to define the imaging parameters to be
used to acquire the projection images. When you progress from this step, the position of the
sample on the manipulator, the X-ray conditions and the imaging device parameters are
recorded.
Corrections
This tab contains the settings used to define and select a shading correction and to set a
safe sample position, which is used when shading correction images are collected. Flux
normalisation and latency correction can also be applied to the X-ray projection images.
Reconstruction
This tab lets you specify whether the images are to be reconstructed by CT Agent after
acquisition and to set the parameters used for the reconstruction. If you want to view and
manipulate the dataset in CT Pro before reconstruction, ensure that you do not choose to
automatically reconstruct the dataset.
Содержание XT V 160
Страница 1: ...X Tek X ray and CT Inspection XT V 160 X ray System Operator Manual XTM0011 F2 ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 22: ......
Страница 26: ......
Страница 32: ......
Страница 83: ...Maintenance XT V 160 X ray System XTM0011 F2 73 ...
Страница 90: ...Troubleshooting 80 XTM0011 F2 XT V 160 X ray System ...
Страница 93: ......