VI Maintenance Menu
4 Initializing Focus Motion Potentiometer
- 63 -
VI
4
Initializing Focus Motion Potentiometer
The potentiometer can be initialized at each reference point to detect absolute position of the focusing mechanism.
Use this function after changing the combination of the microscope and HUB controller.
Initialization takes about 3 minutes to be completed. During initialization, never touch the coarse or fine focus
knobs.
1
Remove the rear panel of the remote control pad and set
DIP switch No.8 to ON.
2
Turn on the power to the HUB controller while holding
down the MODE key.
After the MAINTENANCE MENU screen appears, release
the MODE key.
Select “TEST 2” using the
S
key or the
T
key and press
the [
] key.
MAINTENANCE MENU
1.Version
2.TEST1
3.
TEST2
4.PAD Data
5.HUB Data
6.Control Data
a.All Control Data
b.Revoiving Nosepiece
c.Condenser Cassette
d.Light Path
e.Filter Block
f.Exciter Filter
g.Barrier Filter
h.AUX Filter
3
After the TEST 2 screen appears, press the Z-AXI. RESET
key.
Initialization of the focus motion potentiometer begins.
The initialization takes about 3 minutes at maximum.
4
When the initialization is completed, a “Complete!”
message appears and remains displayed for about one
second.
Also, the HUB controller sounds the buzzer.
Note:
The buzzer sounds as follows:
Upon normal completion:
a peep
Upon failure of normal completion: four peepings
5
Turn off the power to the HUB controller.
Set DIP switch No.8 to OFF and reattach the rear panel to
the remote control pad.
Use the coarse focus knob to lower the focusing
mechanism 2 to 3 mm, then turn on the HUB controller.
Turning the coarse focus knob halfway lowers the focusing
mechanism about 2.5 mm.
Note:
After initialization, the focusing mechanism is
located at the highest position.