Chapter 4
Functions and Operations of the Devices
58
14.3
P-IER Eyelevel Riser
Attach the P-IER between the zooming body and the tube to raise
the eyelevel 25 mm.
Up to two eyelevel risers can be used by putting one on top of the
other.
P-IER Eyelevel Riser
Note on use of the eyelevel riser
Do not use more than two eyelevel risers at once or together with other intermediate tube such as a beam splitter or a coaxial
episcopic illuminator. Doing so will cause vignetting around the image.
14.4
P-THSS Teaching Head
Observation by main and secondary observers is enabled by attaching a zooming body and a tube.
Simultaneous observation of the same viewfield is possible through the eyepieces on the secondary observer side and the
eyepieces on the main observer side, making it ideal for educational purposes.
For details, refer to the instruction manual supplied with the P-THSS Teaching Head.
14.5
P-IDT Drawing Tube
The image of the specimen being observed can be sketched by tracing the image of the specimen.
Both the microscopic image and the sketched image can be seen simulteneously through the binocular tube.
For details, refer to the instruction manual supplied with the P-IDT Drawing Tube.
Содержание SMZ1270
Страница 1: ...Stereo Microscope SMZ800N SMZ1270 SMZ1270i Instruction Manual M609 E 18 5 NF 7 1 2 M609EN07 ...
Страница 2: ......
Страница 18: ...Chapter 1 Components 4 ...
Страница 22: ...Chapter 2 Microscopy Procedures 8 ...
Страница 96: ...Chapter 5 Assembly 82 ...
Страница 99: ...85 Chapter 7 Maintenance and Storage This chapter describes how to maintain and store the product ...
Страница 102: ...Chapter 7 Maintenance and Storage 88 ...
Страница 103: ...89 Chapter 8 Specifications ...