Chapter 4
Functions and Operations of the Devices
57
14
Other Devices
In addition to the devices described in the earlier sections, the following optional devices can be used with this microscope
system:
14.1
SMZ10A Aperture Unit
To increase the depth of focus during image capture, attach this
unit between the zooming body and the tube.
When using this unit with a coaxial epi illuminator, first attach this
unit and then the coaxial epi illuminator above the unit.
Move the aperture diaphragm open/close lever to adjust the
diaphragm appropriately. The
○
marks above the lever
represent the diaphragm diameter sizes that vary between 16 mm
dia. and 6.5 mm dia.
SMZ10A Aperture Unit
14.2
P-IBSS2 Beam Splitter
Attach the P-IBSS2 between the zooming body and the tube to
split the optical path.
Mounting a digital camera on the beam splitter port via a camera
adapter enables image capture of the specimen being observed.
Also read the instruction manual supplied with the digital camera
in use.
P-IBSS2 Beam Splitter
Optical path switching
Switch the optical path by sliding the optical path switching knob.
BINO position
Left eyepiece:Right eyepiece:Port (camera)=100:100:0
PHOTO/BINO position
Left eyepiece:Right eyepiece:Port (camera)=50:100:50 (port positioned on the left)
Left eyepiece:Right eyepiece:Port (camera)=100:50:50 (port positioned on the right)
Port
Optical path
switching knob
100%
100%
50% 50%
100%
BINO
PHOTO/BINO
Содержание SMZ1270
Страница 1: ...Stereo Microscope SMZ800N SMZ1270 SMZ1270i Instruction Manual M609 E 18 5 NF 7 1 2 M609EN07 ...
Страница 2: ......
Страница 18: ...Chapter 1 Components 4 ...
Страница 22: ...Chapter 2 Microscopy Procedures 8 ...
Страница 96: ...Chapter 5 Assembly 82 ...
Страница 99: ...85 Chapter 7 Maintenance and Storage This chapter describes how to maintain and store the product ...
Страница 102: ...Chapter 7 Maintenance and Storage 88 ...
Страница 103: ...89 Chapter 8 Specifications ...