I. Parts of the microscope and their names
9
Note:
Be sure to cover unused ports with supplied caps to prevent the entry of external light and dust.
Observation port
Install the eyepiece tube.
Eyepiece-tube clamp screw
Front port
Photo-mask dial
Clockwise turn: OUT
Counterclockwise turn: IN
Intermediate magnification dial
1X: No intermediate magnification
1.5X: 1.5X intermediate magnification for all ports
Coarse focus knob
Fine focus knob
Adapter clamp screw
Right-side port
Optical path switch-over dial
1: Observation 100%
2: Observation 20%, Right 80%
3: Not used (Bottom port 100%)
4: Observation 20%, Front 80%
5: Left 100%
Objective refocusing ring
Forward turn: Release
Backward turn: Lock
TE2000-U, right side
TE2000-U, left side
Dia-illumination ON/OFF switch
Brightness adjustment dial
This dial is disabled
when the remote control pad is used
for bright adjustment.
Stage
Objective
Mount for cassette holder
For the installation of the cassette holder
for epi-fl microscopy
Left-side port
Black cover for elevating section
Remove the cover when installing
the epi-fl attachment.
Mount for analyzer
Install the analyzer here
for DIC microscopy.
Nosepiece
Fine focus knob
Coarse focus knob
Coarse focus torque adjustment ring
Forward turn: Decreases torque
Backward turn: Increases torque
Adapter clamp screw
The bottom port changeover lever is provided on the left-hand side of the TE2000-U bottom port type.
See P. 41 for optical path switching.