Attaching the Polarizer for the Diascopic Illumination
There are two types of polarizers for the diascopic illumination. One is the C-SP polarizer for the
diascopic illumination equipped with a polarizer only. And another is the D-DP rotatable polarizer
for the diascopic illumination equipped with a polarizer and a 1/4 lambda plate. The C-SP polarizer
is used for the polarization microscopy under the diascopic illumination. The D-DP rotatable
polarizer is used for the polarization DIC microscopy under the diascopic illumination. Select the
polarizer suitable for the microscopy.
1
Check the orientation indicator for the C-SP
polarizer. Or check the index for the D-DP
rotatable polarizer.
2
Place the polarizer onto the field lens on the base
of the microscope. The orientation indicator or the
index must face the front of the microscope.
3
Tighten the setscrew to fix the polarizer. The
orientation of the polarizer must be adjusted for
each observation.
F.S.
7 5 3
7 0 1
JAPA
N
Polarizer for
the diascopic
illumination
7 5 3 7 0 1
JAPAN
Swing-out
center
Orientation indicator
Setscrew
C-SP polarizer for the diascopic
illumination
7 5 3 7 0 1
JAPAN
D - D P
Swing-out
center
Thumbscrew
Orientation
indicator
Index
Setscrew
D-DP rotatable polarizer for the
diascopic illumination
Adjusting the orientation of the polarizer
After attaching the polarizer for the diascopic illumination, adjust the orientation of the
polarizer to a right angle to the orientation of the analyzer. For the procedure of adjusting
the orientation, see “15. Polarizer for the Diascopic Illumination” in “III. Operation of
Each Part.”