LS-50
3. Scanning
・
Scanning misalignment:
Body only (except adapters)
Main scannig direction: ± 0.35 mm or less
・
Sub-scanning direction: ± 0.40 mm or less
MA-21
Main scannig direction: ± 0.7 mm or less
・
Sub-scanning direction: ± 0.5 mm or less
IA-20(S)
Main scannig direction: ± 0.55 mm or less
・
Sub-scanning direction: ± 0.90 mm or less
SA-21
Main scannig direction: ± 0.70 mm or less
・
Sub-scanning direction: ± 1.26-mm or less
・
Trimming misalignment:
1 pixel or less
・
Main scanning magnification:
1.254 times ± 1.83% or less
・
Horizontal to vertical ratio (Aspect ratio):
36.87 (Sub)×25.17 (Main) ± 2% or less
・
Main/sub-scan perpendicularity:
90° ± 0.35°
・
Sub-scan parallelism:
Total area
6 pixels or less within the range of 36.87 mm.
Between adjacent pixels
1 pixel or less for any adjacent pixels in the sub-scanning direction.
Small area
4 pixels or less within any continuous 210 lines in the sub-scanning direction.