5(13)
Servodan A/S | Tel +45 7442 4726 | Fax + 45 7442 4035 | [email protected] | www.servodan.com
GB
matically. This setting should be used when the desired minimum
lighting in the room has been reached (the time when the light is
to be switched on and when there is no longer sufficient daylight
in the room).
Time buttons:
Setting the off-delay time. There are seven preset settings:
Pulse/ 2 min./5 min./10 min./15 min./ 30 min. and 60 min.
There is also an option for eight-hour constant-on for running
in e.g. an HVAC system. The red LED flashes 0.25 sec. On and
30 sec. off when the function is active. To cancel the function,
activate the button again and the red LED stops flashing. The de-
tector also exits from the eight-hour timer, returning to automatic;
the light can now be switched on according to the selected
application. When “8 Hours” is active, all other functions are
blocked on the detector.
Instead of selecting a normal delay, it is possible to choose the
“Pulse function”, where the output pulses, i.e. flashes with a
fixed rhythm. Pulse is 5 seconds On and 55 seconds Off. The
pulse function can be used e.g. to control a ventilation system.
The pulse function stops after the selected off-delay time, after
the last activity detection in the sensor’s coverage area has
ceased. If this function is desired, first press “Pulse”, then press
one of the “Time” buttons.
Example:
The “Pulse” and “5 min” buttons are activated. This means that
the detector will pulse 5 seconds On and 55 seconds Off when
there is activity in the detector’s coverage area. The detector
discontinues the pulses 5 minutes after the last registered activ-
ity.
Fig. 2.
In order to cancel the pulse function, choose one of the other
“Time” buttons, without first activating the “Pulse” button.
Status:
If this button is activated, the detector will indicate which set-
tings it is operating from via the red LED.
The green LED indicates start status.
Example:
The detector is set to: Lux value = 300 Lux and Time = 10 min.
If the “Status” button is pressed, the green LED will flash once,
the red LED will flash five times, followed by one flash from the
green LED and then four red flashes. The first five red flashes
indicate the fifth lux setting = 300 lux. The following four red
flashes indicate the fourth Time setting = 10 minutes.
Lux
20 50 100 200 300 400 500 700 1000
No
lux
Learn
Number of
flashes
1
2
3
4
5
6
7
8
9
10
11
Time
Pulse
2
minutes
5
minutes
10
minutes
15
minutes
30
minutes
60
minutes
Number of
flashes
1
2
3
4
5
6
7
Test On/Off:
Activating walk-test.
When the test has been activated, the red LED comes on. When
the detector detects activity in the coverage area, the red LED
will flash and the associated lighting will switch on for 5 seconds.
NB! The walk-test is daylight-
independent
.
Short push:
Has the same function (and limitations) as a short push given by
the operating button.
Long push:
Has the same function (and limitations) as a long push given by
the operating button.
Fact. setting:
Pressing this button immediately restores all settings to factory
settings.
Memory in:
Saves settings to the remote control’s memory (most recently
entered Lux value and most recently entered Time value).
Memory out:
Sends the saved settings to the sensor.
Clear Memory:
Deletes settings from the remote control’s memory.
Summary of the LED flash signals:
Status
Indication
Start up
The red LED flashes 0.25 sec. On and 2 sec. Off.
Concludes with two short flashes from the green LED.
The warm-up time is approximately 40 seconds. That
relay is active during warm-up.
“Long press”
mode
The red LED flashes 0.25 sec. On and 5 sec. Off.
“Light On/Off” The red LED flashes 0.25 sec. On and 10 sec. Off.
Error, activa-
tion
of 230 V but-
ton
The green LED and red LED flash 0.25 sec. On and 5
sec. Off in alternation. This either means that the 230
V button was not correctly activated, or that the mo-
tion detector has not correctly received a signal from
the 230 V button.
“8 Hour”
mode
The red LED flashes 0.25 sec. On and 30 sec. off
when the function is active.