background image

ProAxis

plus

™  Electric Profiling Bed and Pressure Area Care System  

Instruction for use and trouble shooting guide 

________________________________________________________________________________ 

ProAxis Plus Operating Instructions 3.1                Dated 18/05/11                                           12/24 

2.6 Operating Instructions 

 
Before using ProAxis

plus

, ensure the Operations and Safety Installation section has been 

followed. Important: ProAxis

plus

 should be plugged into the mains at all times, the battery 

backups are for emergencies only. 
 
Positioning the Patient 
 
Use the patient / nurse handset to slightly profile the bed to identify the dropping hinge 
section. Ensure the sacrum area is above the hinge between the back-rest and thigh 
section of the platform.  
 
This will ensure that when the backrest is raised, the  hinge drops, the automatic knee-
break is operated, and the patient does not slip down the mattress.   
 
In normal circumstances, it will be necessary to use no more than two pillows. If the patient 
requires sitting up the backrest should be raised and the use of pillows for “propping” the 
patient up should be avoided. 
 
Safety Sides & Headboards 
 
The system will normally be supplied with safety sides fitted and positioned either side of 
the mattress platform. These are adjusted by pushing in the black button on the side rail 
bracket. Lift or drop until they click into position. The safety sides should only be used after 
full assessment of the patient has been completed, as their use may not be suitable in all 
cases.  
 
It is essential that the head and footboards are fitted when the system is to be tilted.  
 
Braking System 
 
The brakes have directional steer 
and 4 way locking. 

 

Directional 
 

 
No Lock 
 
 

4 Way Lock 

 

 

 

The braking system is operated by pedals located at the foot end of the bed. When the 
pedal is horizontal the brakes are off. With the pedal depressed towards the foot end the 
brakes are locked on. With the pedal depressed towards the head end the steering lock is 
activated to assist in moving the bed.

 

 
For pressure relief, use the Smart Control (optional. See Smart Control Operating 
Instructions for instructions of use. 
 

Содержание ProAxis Plus

Страница 1: ...ea Care System Instruction for use and trouble shooting guide ________________________________________________________________________________ ProAxis Plus Operating Instructions 3 1 Dated 18 05 11 1...

Страница 2: ...uk Website www nightingalebeds co uk Nightingale Care Beds Ltd declares that the ProAxisplus Electric Profiling Bed and Pressure Area Care System have been designed and manufactured to meet the follo...

Страница 3: ...Profiling Bed Pressure Area System 1 1 Introducing ProAxisplus 1 2 Achievable Positions 1 3 The Mattress 2 Operations and Safety 2 1 Installing the ProAxisplus 2 2 Safety Precautions 2 3 ProAxisplus...

Страница 4: ...ed pressure area care system that profiles to the contours of the human body by incorporating an automatic knee break whilst dropping the sacral area This unique profiling movement alleviates three ex...

Страница 5: ..._____________________ ProAxis Plus Operating Instructions 3 1 Dated 18 05 11 5 24 1 2 Achievable Positions ProAxisplus can achieve many positions by easy operation of a simple handset Every function i...

Страница 6: ...nd helps to dissipate moisture and perspiration away from the body It is common for modern mattresses to have anti bacterial covers but the Nightingale mattress now achieves the same standard internal...

Страница 7: ...al should be pressed down into the lock position before any attempt is made to place a patient onto ProAxisplus Before placing the patient onto ProAxisplus the mattress platform should be profiled sli...

Страница 8: ...ufficient clearance between the frame and the wall Care should be taken when the system is being moved and the power supply cable must be unplugged and stored with the system whilst in transit Battery...

Страница 9: ...________ ProAxis Plus Operating Instructions 3 1 Dated 18 05 11 9 24 Operating Instructions in Diagrammatic Format 2 3 Main Features Backrest Removable Head Foot Boards Safety Buffers Dropping Hinge C...

Страница 10: ...own Backrest To raise or lower the backrest of the platform the buttons marked Head Up and Head Down should be used Footrest To raise or lower the footrest the buttons marked Foot Up and Foot Down sho...

Страница 11: ...e stopped anytime after two seconds into the movement by pressing ONE of the red CPR buttons once Tilt The system can be tilted by pressing one of the two green buttons on the lockout box marked Tilt...

Страница 12: ...ws If the patient requires sitting up the backrest should be raised and the use of pillows for propping the patient up should be avoided Safety Sides Headboards The system will normally be supplied wi...

Страница 13: ...ans that when the head section moves the thigh and foot sections move also Combining this movement with the proven capabilities of the Visco Elastic mattress gives the opportunity to offer a high degr...

Страница 14: ...______________________________________________________ ProAxis Plus Operating Instructions 3 1 Dated 18 05 11 14 24 Smart Operation Screen Identification Home Screen Press BLUE function key below Smar...

Страница 15: ...d number is achieved press ACCEPT key CONFIRM the settings and if correct press ACCEPT key If the settings are wrong press the REJECT key and start again Smart Settings From the Home Screen press SMAR...

Страница 16: ...operating however the CPR function will still be available Once activated the CPR function will disable Smart until the programme is reactivated 5 Please ensure that the patient is able to tolerate mo...

Страница 17: ...t exceed the safe working load of any equipment utilised to prevent manual handling issues The system works by calculating the enery required to lift the patient up and down as this is directly propor...

Страница 18: ...ing Bed and Pressure Area Care System Instruction for use and trouble shooting guide ________________________________________________________________________________ ProAxis Plus Operating Instruction...

Страница 19: ...viruses Nightingale treat blood or body fluid spills with NaDCC solution at 10 000 parts per million available chlorine Routing Decontamination NaDCC solution at 1000 parts per million available chlo...

Страница 20: ...__________ ProAxis Plus Operating Instructions 3 1 Dated 18 05 11 20 24 All of the above Important If there is any doubt as to what solution of cleaning agent is required for the various cleaning proc...

Страница 21: ...em is supplied with a 3 year warranty subject to servicing annually covering labour and parts that may require replacing through normal use It is recommended that every system is serviced annually to...

Страница 22: ...Foot Rest Angle 0 17 Degrees below electric variable Knee Break Electric Auto Variable CPR Electric Auto Mass 165kg Capacity 318kg Safe Working Load 353kg Trendelenburg Tilt 12 5 Degrees Reverse Tren...

Страница 23: ...ed by a red screen If in Smart Weigh Mode press exit then press manual If the confirm settings screen is displayed press manual Transformer Batteries Is the transformer attached plugged into the wall...

Страница 24: ...______________________________ ProAxis Plus Operating Instructions 3 1 Dated 18 05 11 24 24 Total after sales service Our aftersales and Helpline are always available Nightingale Care Beds Ltd Unit 1...

Отзывы: