60
www.nieaf-smitt.com
Nieaf-Smitt is a brand name of Mors Smitt
6
Specifi cations
General specifi cations
Display
40.000 count backlit liquid crystal with
bargraph
Polarity
Automatic (no indication for positive);
minus (-) sign for negative
Measurement rate
2 times per second, nominal
Overrange indication
‘OL’ is displayed
Battery
6x 1.5 V AA battery
Low battery indication
is displayed if battery voltage drops
below operating voltage
Auto power off
Approx. 15 minutes (with disable
function)
Fuses
mA, µA ranges; 0.5 A/1000 V ceramic
fast blow A range; 10 A/1000 V ceramic
fast blow
Operating temperature +5 ºC...+40 ºC (+41 ºF...+104 ºF)
Storage temperature
-20 ºC...+60 ºC (-4 ºF...+140 ºF)
Operating humidity
Max 80 % up to 31 ºC (87 ºF) decreasing
linearly to 50 % at 40 ºC (104 ºF)
Storage humidity
< 80 %
Store capacitance
2000
Weight
631 g
Dimensions
220 x 96.5 x 60.5 mm
Diode test
Test current of 0.9 mA maximum, open
circuit voltage 2.8 VDC typical
Continuity check
Audible signal will sound if the resis-
tance is less than 35 Ω (approx.), test
current < 0.35 mA
Peak
Captures peaks >1 ms
Input impedance
>10 MΩ VDC & > 9 MΩ VAC
AC response
True RMS
AC True RMS
The term stands for Root-Mean-Square,
which represents the method of calcu-
lation of the voltage or current value.
Average responding multimeters are
calibrated to read correctly only on sine
waves and they will read inaccurately on
non-sine wave or distorted signals. True
rms meters read accurately on either
type of signal.
ACV bandwidth
50 Hz....1000 Hz
Crest factor
≤ 3 at full scale up to 500 V, decreasing
linearly to ≤ 1.5 at 1000 V