Nidek Medical Mark 5 Nuvo Lite Family Скачать руководство пользователя страница 6

6

May 2011

2010-8400 Rev F

Electrical power supply:

115 V Units

230 V Units

Frequency:

60Hz

50 & 60Hz

Average Power:

330 W (avg)

300 W (avg)

Protection Class:

Class II

Class II

Mains Protection:

10A

5A

Filters:

At the rear of the device: a cabinet air (dust)  lter.

At the compressor input: an inlet air  lter, 5 µm, behind 

the cabinet  lter.   

Before the oxygen outlet: a  nal product  lter <0.3 µm. 

(technician only).

Air circulation:

A tubeaxial fan cools the compressor compartment.
.

Environmental limit conditions:

The performances of the device (especially the oxygen
concentration) are quoted at 21

o

C (70

o

F) and one atmos-

phere. They may change with temperature and altitude.
For further information, please consult the maintenance
manual.
• The device must be stored, transported and used in the
vertical position only.
• Ambient temperature of between 5

o

C and 40

o

C (40

o

F

and 104

o

F) (operation).

• Storage temperature range -20

o

C to 60

o

C. (-4

o

F

to140

o

F).

• Relative humidity of between 15 % and 95 % operation
and less than 95% storage, both non-condensing.
• Altitude(21

o

C): Up to 2,286 m (7,500 ft) without de-

gradation;
Consult your equipment provider for further information
regarding 2,286 m to 4000 m (7,500 to 13000 ft).
• Complies with EN60601-1 standard; spilling of a glass
of water.

IV. 7. Standards

EN 60601-1[UL60601-1:2003],CAN/CSA-C22.2
No.601.1-M90 w/A1&A2: Electrical Safety- Medical
Devices.
EN60601-1-2:2001 Electromagnetic Compatibility

I

V. 8. Method for disposing of waste

All waste from the

MARK 5 Nuvo LITE

(patient circuit,

 lter, etc.) must be disposed of using the methods appropriate 

to the civil authority of the location where disposed.

IV. 9. Method for disposing of the device

In order to preserve the environment, the concentrator
must only be disposed of using the appropriate methods.
All materials of construction are recyclable.

MARK 5 Nuvo LITE

Serial No. ______________

Date  rst used: 

___________________________

_________________________________________

Maintained by: ___________________________
_________________________________________

Your equipment supplier: __________________

Address : ________________________________
_________________________________________
_________________________________________

Telephone : ______________________________

PREVENTIVE MAINTENANCE:

a

. Wash cabinet  lter weekly.

b. Inspect inlet air  lter at each patient visit. 

    Replace  lter every 2 years, or more often 

depending on environment.

c. Check oxygen concentration every 5,000 hours or 1

year.

The manufacturer’s instructions for the

preventive

maintenance

of the devices are defined in the

maintenance manual. Check with your service provider
for any updates to the recommended schedules.

The work must be carried out by suitably trained tech-

nicians certi ed by the manufacturer.

Use original spare parts only (see Pg. 7).

Upon request, the supplier can provide circuit diagrams,
spare parts lists, technical details or any other information

of use to quali ed technical personnel for parts of the 

device which are designated as being the manufacturer’s
responsibility or by the manufacturer as repairable.

Medical Device Regulations require users and
service providers to report to the manufacturer any
incident that could, if repeated, result in injury to
any person.

StockCheck.com

Содержание Mark 5 Nuvo Lite Family

Страница 1: ...6 IV 9 Method for disposing of device 6 IV 10 Troubleshooting 7 APPENDIX A EMC INFORMATION 8 9 CONTENTS ON power switched on Off power switched off Type B device Class II protection Do not expose to o...

Страница 2: ...tor are not considered to be responsible themselves for the consequences on the safety reliability and characteristics of a device unless Theassembly tting extensions adjustments modi cations or repai...

Страница 3: ...ower cable into a power outlet of the correct voltage and frequency as de ned on the manufacturer s technical label Fig I2 10 f Press the power switch I O to the ON position I The red indicator will l...

Страница 4: ...ftheMARK5NuvoLITE istobecleaned with a soft dry cloth or if necessary a damp sponge then thoroughly dried with wipes and an alcohol based solution Acetone solvents or any other in ammable products mus...

Страница 5: ...dels Class II device with insulated casings EN60601 1 stan dard Pressure A safety relief valve calibrated to 2 7 bar 40 psig is tted on the compressor outlet IV 5 3 Maintenance of the system alarms No...

Страница 6: ...dofusingthemethodsappropriate to the civil authority of the location where disposed IV 9 Method for disposing of the device In order to preserve the environment the concentrator must only be disposed...

Страница 7: ...eumatic circuit fault Compressor thermal safety device has been activated Fan is not working Dirty Filters Tubedisconnected orhumidi er not tight Cannula tubing is kinked Observations The I O ON OFF b...

Страница 8: ...8 Jan 2011 2010 8400 Rev F S t o c k C h e c k c o m...

Страница 9: ...9 Jan 2011 2010 8400 Rev F S t o c k C h e c k c o m...

Страница 10: ...__________________________________________ _________________________________________________________________ _________________________________________________________________ _________________________...

Страница 11: ...____________________________________ __________________________________________________________________ __________________________________________________________________ _____________________________...

Страница 12: ...2010 8400 Rev F Nidek Medical Products Inc 3949 Valley East Industrial Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 info nidekmedical com www nidekmedical com S t o c k C h...

Отзывы: