III.3. Maintenance

No special maintenance needs to be carried out by the
patient.

Your equipment supplier performs periodic main-

tenance operations to assure continued reliable service from
the

MARK 5 Nuvo LITE

.

IV. USEFUL INFORMATION

IV.1. Accessories and spare parts

The accessories used with the

MARK 5 Nuvo LITE

must:

• be oxygen compatible,
• be biocompatible,
• comply with the general requirements of the FDA Quality
System Regulation.

The connectors, tubes and nasal cannulas must be designed
and approved for oxygen therapy usage.

The accessories with a

Nidek Medical

part number reference,

or included in the set of accessories supplied with the device,
comply with these requirements.
Contact your equipment provider to obtain these accesso-
ries.

Note:

The use of certain administration accessories which

are not speci ed for use with this concentrator may reduce 

its performance and void the manufacturer’s responsibility.

AVAILABLE ACCESSORIES

IF PRESCRIBED BY A PHYSICIAN

    Humidi er: 

 

 

       P/N  9012-8774

Cannula with 2 m (7 ft) tubing:

P/N 9012-8780

Extension Tubing 7.7 m (25ft):

P/N 9012-8781

Tubing Adapter:

P/N 9012-8783

The above items are available from

Nidek Medical Products, Inc.

IV.2. Materials in direct or indirect contact with the
patient

Concentrator casing.............................. ABS/Polycarbonate
Mains cable ................................................................. PVC

Dust  lter ............................................................. Polyester

ON/OFF switch ......................................................... Nylon
Casters ....................................................................... Nylon
Flow adjustment knob ................................................. ABS
Oxygen Product outlet.......................................Metal Alloy
Printed labels ................................................ Polycarbonate
Pipe/Tubing......Aluminium, PVC,polyurethane or silicone

Humidi er  ................................................... Polypropylene
Humidi er Tubing.........................................................PVC

Filter ............................................................. Polypropylene

6

7

8

9

10

I

II. CLEANING - MAINTENANCE

III.1. Cleaning

Only the outside of the

MARK 5 Nuvo LITE

is to be cleaned,

with a soft, dry cloth or, if necessary, a damp sponge, then
thoroughly dried with wipes and an alcohol based solution.

Acetone, solvents or any other in ammable products 

must

not be used

. Do not use abrasive powders.

The removable Cabinet Air (dust)  lter (2) must be cleaned 

weekly in warm water and household detergent or after
approximately 100 hours of use. More frequent cleaning is re-
commended in dusty enviroments. Dry before reinstalling.

a. The following minimum guidelines should be
observed:

• Humidi er: 

(If prescribed by a physician)

Clean according to the manufacturer's instructions. If no
instructions are provided, do the following:

Daily:

- Empty the water from the humidi er.

- Rinse the humidi er  ask under running water.

- Fill humidi er up to the mark with distilled water.

Regularly:

-  Disinfect  the  humidi er  parts  by  immersing  them  in  a 

disinfectant solution (in general, we recommend using a
solution of 1 part vinegar diluted with 10 parts water).
- Rinse and dry.

- Check that the humidi er lid seal is in good condition.

• Oxygen tubing and nasal cannula:
Follow the manufacturer’s instructions.

b. For each new patient:

Follow the instructions from the humidi er manufacturer.

The

MARK 5 Nuvo LITE

cabinet should be cleaned and

disinfected as per the above instructions. The cabinet air

 lter should be washed or replaced. The entire oxygen ad-

ministration circuit (oxygen therapy nasal cannulas, etc.)
must be changed.

III.2. Daily disinfection

Due to the presence of the  nal product  lter <0.3 µm inside 

the device, everyday disinfection concerns only the external

oxygen therapy accessories: humidi er, probes, nasal can-

nulas (refer to the respective instructions for use).

The device must be switched off when alcohol based
solutions are used.

1. Filter (5 µm)/Silencer

2. Cabinet Air (dust)  lter                           

3. Ventilation grill

7

8

StockCheck.com

Содержание Mark 5 Nuvo Lite Family

Страница 1: ...6 IV 9 Method for disposing of device 6 IV 10 Troubleshooting 7 APPENDIX A EMC INFORMATION 8 9 CONTENTS ON power switched on Off power switched off Type B device Class II protection Do not expose to o...

Страница 2: ...tor are not considered to be responsible themselves for the consequences on the safety reliability and characteristics of a device unless Theassembly tting extensions adjustments modi cations or repai...

Страница 3: ...ower cable into a power outlet of the correct voltage and frequency as de ned on the manufacturer s technical label Fig I2 10 f Press the power switch I O to the ON position I The red indicator will l...

Страница 4: ...ftheMARK5NuvoLITE istobecleaned with a soft dry cloth or if necessary a damp sponge then thoroughly dried with wipes and an alcohol based solution Acetone solvents or any other in ammable products mus...

Страница 5: ...dels Class II device with insulated casings EN60601 1 stan dard Pressure A safety relief valve calibrated to 2 7 bar 40 psig is tted on the compressor outlet IV 5 3 Maintenance of the system alarms No...

Страница 6: ...dofusingthemethodsappropriate to the civil authority of the location where disposed IV 9 Method for disposing of the device In order to preserve the environment the concentrator must only be disposed...

Страница 7: ...eumatic circuit fault Compressor thermal safety device has been activated Fan is not working Dirty Filters Tubedisconnected orhumidi er not tight Cannula tubing is kinked Observations The I O ON OFF b...

Страница 8: ...8 Jan 2011 2010 8400 Rev F S t o c k C h e c k c o m...

Страница 9: ...9 Jan 2011 2010 8400 Rev F S t o c k C h e c k c o m...

Страница 10: ...__________________________________________ _________________________________________________________________ _________________________________________________________________ _________________________...

Страница 11: ...____________________________________ __________________________________________________________________ __________________________________________________________________ _____________________________...

Страница 12: ...2010 8400 Rev F Nidek Medical Products Inc 3949 Valley East Industrial Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 info nidekmedical com www nidekmedical com S t o c k C h...

Отзывы: