IV. 10. Troubleshooting.

   

           

Maintenance Items

 

Cabinet Air Filter: 

Part Ref: 9250-1025;   Wash weekly; Replace as needed.

 

Inlet Air Filter:    

Part Ref:  9250-1180;  Inspect at each patient visit; Replace annually.   

 

 

 

Please  record  all  maintenance  activity  in  the  Maintenance  Log  found  in  the  service  manual  and  online  at                      

www.nidekmedical.com under the 'Maintenance Log' tab.  

Observations

       

Possible Causes

                    Solutions    

The

 I-0

 (ON/OFF) button is in the

 

“I”

 (ON) position but the device 

does not operate.

Power cable (9) is not correctly 

plugged into the wall outlet.

Check the cable connection.

The audible alarm sounds intermit

-

tantly.

Power failure.

Check the circuit breaker (5) on the 

front of the unit; Reset if necessary.

Red light remains lighted.

Oxygen concentration is too low. 

or pressures are out of specified 

range.

Contact your equipment supplier.

The alarm test does not work. See 

IV 4.1.

Capacitor  is  charged                                                                                                                                          

                                                               

Internal electrical fault.

Plug unit in for 10 minutes and retest

Contact your equipment supplier.

The compressor operates and the 

I-0

 

(ON/OFF) button is in the 

 “I”

 (ON) 

position but the green indicator is 

not lighted.

Faulty indicator.

Contact your equipment supplier.

The

 I-0

 (ON/OFF) button is in the

 

“I”

 (ON) position but there is no 

flow. The audible alarm sounds con-

tinuously.

Pneumatic connection broken or 

other pressure problem.

Stop the device by pressing the 

I-0

 

(ON/OFF) button and contact your 

equipment supplier.

The

 I-0

 (ON/OFF) button is in the

 

“I”

 (ON) position, the compressor is 

operating and there is a flow but the 

audible alarm sounds continuously.

Internal electrical fault.

Pneumatic circuit fault or low 

purity.

Stop the device and contact your 

equipment supplier.

The compressor stops in mid-cycle, 

then starts again after a few minutes.

Compressor thermal safety device 

has been activated.

Dirty Filters.

Fan is not working.

Stop the device and wait for it to cool 

down.

Clean cabinet filter.  Restart.

If the device does not start, contact 

your equipment supplier.

The oxygen enriched air flow is 

interrupted at the nasal cannula 

outlet.

Tube disconnected or humidifier 

cap is not tight. 

Check that tubing connections are se

-

cure and that the humidifier is sealed.

The flow at the nasal cannula outlet 

is irregular.

Cannula tubing is kinked or  res

-

tricted.

Straighten the tubing; contact your 

equipment supplier if damaged.

Содержание Mark 5 Nuvo 10

Страница 1: ...3 III CLEANING MAINTENANCE 4 III 1 Cleaning 4 III 2 Everyday disinfection 4 III 3 Maintenance 4 IV USEFUL INFORMATION 4 IV 1 Accessories and spare parts 4 IV 2 Materials in direct indirect contact wi...

Страница 2: ...r installer or distributor are not considered to be responsible themselves for the consequences on the safety reliability and characteristics of a device unless Theassembly fitting extensions adjustme...

Страница 3: ...of the correct voltage and frequency as defined on the manufacturer s technical label 8 f Press the power switch I 0 to the ON position I The green indicator will light when the oxygen concentration e...

Страница 4: ...ons The cabinet air filter should be washed or replaced The entire oxygen administration circuit oxygen therapy nasal cannulas etc must be changed III 3 Maintenance No special maintenance needs to be...

Страница 5: ...angle transport with humidifier fitted 70o Weight 24 kg 54 lbs Noise level within ISO 8359 guidelines Flow values Continuously Adjustable Flowmeter 1 to 5 2 to 8 or 2 to 10 depending on model IV 5 3...

Страница 6: ...the maximum recommended flow does not exceed 10 of the indicated value when a back pressure of 6 9 kPa 1 psig is applied to the output of the device The maximum outlet pressure is 7psig for 5 l min 15...

Страница 7: ...vice provider for any updates to recommended schedules The work must be carried out by suitably trained technicians certified by the manufacturer Use original spare parts only see Pg 8 Uponrequest the...

Страница 8: ...in the I ON position but the green indicator is not lighted Faulty indicator Contact your equipment supplier The I 0 ON OFF button is in the I ON position but there is no flow The audible alarm sound...

Страница 9: ...November 2013 2010 2324CE Rev A 9 V EMC Electromagnetic Statements...

Страница 10: ...010 2324CE Rev A 10 November 2013 NOTES...

Страница 11: ...November 2013 2010 2324CE Rev A 11 NOTES...

Страница 12: ...alley East Industrial Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 EU Representative mdi Europa GmbH Langenhagener Str 71 30855 Hannover Langenhagen Germany Tel 49 511 39 08...

Отзывы: