III. CLEANING - MAINTENANCE

III.1. Cleaning

Only the outside of the 

Mark 5 Nuvo MB

 is to be cleaned, 

with a soft, dry cloth or, if necessary, a damp sponge, then 

thoroughly dried with wipes and an alcohol based solution. 

Acetone, solvents or any other inflammable products 

must 

not be used

.  Do not use abrasive powders.

The removable cabinet air filter (7) must be cleaned in warm 

water and household detergent weekly or after approximately 

100 hours of use. More frequent cleaning is recommended in 

dusty enviroments. Dry before reinstalling.  

 

 

                                  

66

 

9

.

II.2. Daily disinfection

Because there is a final product filter inside the device, daily 

disinfection concerns only the external oxygen therapy 

accessories: humidifier, probes, nasal cannulas (refer to the 

respective instructions for use).

The device must be switched off when alcohol based so-

lutions are used.

a.  The following minimum guidelines must be 

observed: 

 Humidifier : (If prescribed by a physician)

Clean according to the manufacturer's instructions.  If no 

instructions are provided, do the following:

Daily:

• Empty the water from the humidifier.

• Rinse the humidifier flask under running water.

• Fill humidifier up to the mark with distilled water.

Regularly:

•  Disinfect  the  humidifier  parts  by  immersing  them  in  a 

disinfectant solution (In general, we recommend using a 

solution of 1 part vinegar diluted with 10 parts water).

• Rinse and dry.

• Check that the humidifier lid seal is in good condition.

 Oxygen tubing and nasal cannula:

Follow the manufacturer’s instructions.

b. For each new patient:

Follow the instructions from the humidifier manufacturer.

The 

Mark 5 Nuvo MB 

must be cleaned and disinfected as 

per the above instructions. The cabinet air  filter should be 

washed or replaced.  The entire oxygen administration circuit 

(oxygen therapy nasal cannulas, etc.) must be changed.

                        

            

III.3. Maintenance

No special maintenance needs to be carried out by the 

patient

. Your equipment supplier performs periodic main

-

tenance operations to assure continued reliable service from 

the 

Mark 5 Nuvo MB

.

IV. USEFUL INFORMATION

IV.1. Accessories and spare parts

The accessories used with the 

Mark 5 Nuvo MB

 must:

  be oxygen compatible,

  be biocompatible,

  comply with the general requirements of the FDA  

   Quality System Regulation or the 93/42/EEC European  

  Directive as appropriate.

 

The connectors, tubes, nasal cannulas, probes or masks    

must be designed for oxygen therapy usage.

The accessories with a 

Nidek Medical

 part number

 

 

reference, 

or included in the set of accessories supplied with the device, 

comply with these requirements.

Contact your dealer to obtain these accessories.

Note: 

 The use of certain administration accessories which 

are not specified for use with this concentrator may reduce 

its performance and void the manufacturer’s responsibility 

(ISO 8359).

AVAILABLE ACCESSORIES

IF PRESCRIBED BY A PHYSICIAN

  Humidifier: 

 

P/N 

9012-8774

  Humidifier High Flow: 

 

P/N 

9251-8774

  Cannula with 2 m (7 ft) tubing: 

P/N 

9012-8780

  Cannula High Flow: 

             P/N 

9251-8780

  Extension Tubing 7.7 m (25ft): 

P/N 

9012-8781

  Tubing Adapter: 

 

P/N   9012-8783

 

   The items listed above are available from 

 

            Nidek Medical Products, Inc.

IV.2. Materials in direct or indirect contact with the 

patient

Concentrator casing................................ABS/Polycarbonate

Mains cable ................................................................... PVC

Cabinet air filter

 ..................................................... Polyester

I/0 (On/Off) switch

...................................................... Nylon 

Casters ......................................................................... Nylon

Flow adjustment knob....................

 ...............................

ABS

Oxygen product outlet  ................................................. Brass

Printed labels .................................................. Polycarbonate

Pipe/Tubing........Aluminium,PVC, polyurethane or silicone

Humidifier 

 .....................................................Polypropylene

Filter  ..............................................................Polypropylene

 

             

6  Filter / Silencer

 7    Cabinet Air Filter

 8    Hour Meter 

 9    Ventilation Grille

 

7

8

6

9

10

8

7

6

Содержание Mark 5 Nuvo 10

Страница 1: ...3 III CLEANING MAINTENANCE 4 III 1 Cleaning 4 III 2 Everyday disinfection 4 III 3 Maintenance 4 IV USEFUL INFORMATION 4 IV 1 Accessories and spare parts 4 IV 2 Materials in direct indirect contact wi...

Страница 2: ...r installer or distributor are not considered to be responsible themselves for the consequences on the safety reliability and characteristics of a device unless Theassembly fitting extensions adjustme...

Страница 3: ...of the correct voltage and frequency as defined on the manufacturer s technical label 8 f Press the power switch I 0 to the ON position I The green indicator will light when the oxygen concentration e...

Страница 4: ...ons The cabinet air filter should be washed or replaced The entire oxygen administration circuit oxygen therapy nasal cannulas etc must be changed III 3 Maintenance No special maintenance needs to be...

Страница 5: ...angle transport with humidifier fitted 70o Weight 24 kg 54 lbs Noise level within ISO 8359 guidelines Flow values Continuously Adjustable Flowmeter 1 to 5 2 to 8 or 2 to 10 depending on model IV 5 3...

Страница 6: ...the maximum recommended flow does not exceed 10 of the indicated value when a back pressure of 6 9 kPa 1 psig is applied to the output of the device The maximum outlet pressure is 7psig for 5 l min 15...

Страница 7: ...vice provider for any updates to recommended schedules The work must be carried out by suitably trained technicians certified by the manufacturer Use original spare parts only see Pg 8 Uponrequest the...

Страница 8: ...in the I ON position but the green indicator is not lighted Faulty indicator Contact your equipment supplier The I 0 ON OFF button is in the I ON position but there is no flow The audible alarm sound...

Страница 9: ...November 2013 2010 2324CE Rev A 9 V EMC Electromagnetic Statements...

Страница 10: ...010 2324CE Rev A 10 November 2013 NOTES...

Страница 11: ...November 2013 2010 2324CE Rev A 11 NOTES...

Страница 12: ...alley East Industrial Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 EU Representative mdi Europa GmbH Langenhagener Str 71 30855 Hannover Langenhagen Germany Tel 49 511 39 08...

Отзывы: