61
CHECK AND ADJUSTMENT PROCEDURES: Groove Position Check and Adjustment
3.12 Groove Position Check and Adjustment
Move the groove position of the auto-processed lens to the desired position.
3.12.1 Groove position check
1
Enter service mode from the LE-700 home screen.
“3.2.1 Entering service mode” (page 20)
2
Press the “Maintenance” tab.
The maintenance screen appears.
3
Press Circle 45
.
The processing condition screen appears.
4
Specify the grooving conditions.
1)
Set each item as follows:
“Material”
→
CR39
“Type”
→
Single
“Frame”
→
Groove
“Groove position”
→
Auto
“Soft”
→
None
“Polish”
→
None
“Layout”
→
Passive
“FC”
→
None
2)
Press .
The layout screen appears.
5
Continue the processing method on the layout screen using the same procedures
explained by the following steps starting from size setting.
to
CAUTION
• The following procedure is to change the position of an auto-processed groove according to the
user’s needs. Do not perform this procedure to correct deviation of the bevel position from the spec-
ified settings. If such an adjustment is found to be required during installation, refer to Service Man-
ual for further details.