57
CHECK AND ADJUSTMENT PROCEDURES: Polish Size Check and Adjustment
2)
Press .
The layout screen appears.
5
Continue the processing method on the layout screen using the same procedures
explained by the following steps starting from size setting.
to
6
Measure the outer diameter of the processed lens with a caliper to check the value is
in the following tolerance.
•
For beveling:
φ
45.0±0.1 mm
•
For flat edging:
φ
45.0±0.1 mm
7
When the lens outer diameter is out of the tolerance of
φ
45.0 ± 0.1 mm, change the
Polish size diameter.
“3.11.2 Polish size adjustment” (page 58)
±
PP
±
PP