If the period of the external input is less than the setting of the 
delay time, the delay time is ignored and the ST-329 flashes as if 
the delayed time equals 0.

To adjust the delay time, press the “+” or “-” key. Or turn the dial 
to set the delay time. Delay time will increase as the dial is turned 
clockwise, and decrease as it is turned counter clockwise.

Delay Angle Setting
As the period of the external input is 360, the delay angle can be 
set from 0°to 360°, by every 3°. Since the internal calculation time 
is 60 µs, the actual delay time is as follows:

Delay angle setting/360° x period of external input + appox. 60 µs.
The ST-329 does not flash at the 1st trigger pulse as shown below 
in the diagram. 
Example: Trigger: positive     Delay Angle: 36°     36/360=10

If the current period of external trigger input changes, the time of 
the flash is not accurate, because the time is calculated based on 
the previous measurement period.

If the current period of external trigger input is less than the previ-
ous period and the next trigger input occurs before the flash time, 
the delay angle setting is ignored and the ST-329 flashes at the 
delay angle=0°.

To adjust, press the “+” or “-” key. You may also use the adjust-
ment dial.
 
The delay angle increases as the dial is rotated to the right. The 
angle settings will go to 0° as the angle increases past 359°.

Delay angle decreases as the dial is rotated to the left. The set-
tings will eventually go to 0° as the angle decreases past 359°.

Parameter Settings
To enter parameter setting menu, press “MODE” until the LCD dis-
plays “P”. When the display shows “P’, press the SET key to cycle 
between the various parameter settings available (range, trigger 
edge, delay time, LCD backlight, Auto Shut-off, Initial). To store 
the settings and return to measuring modes, press the MODE key.

Measuring Range Setting
You can adjust the flash setting from the home view and also in 
parameter mode if desired.

Trigger Edge Setting (External Modes)
In external mode, the trigger edge parameter will allow for select-
ing either the leading edge or trailing edge of the input pulse to 
trigger the stroboscope flash.

In parameter mode, use the SET key to cycle to the Trigger Edge 
setting parameter mode. To set the trigger edge as “Down Edge” 
(trailing), press “ -” key or turn the dial clockwise. To set the trig-
ger edge as “Up Edge” (leading), press “+” key or turn the dial 
clockwise.

When trigger edge setting is set to “up edge”. The flash will oc-
cur on the leading edge of the input pulse (when delay setting is 
zero).

External Pulse Signal from Machine/Sensor 

vs. Stroboscope Flash

Up Edge

When trigger edge setting is set to “down edge”. The flash will 
occur on the trailing edge of the input pulse (when delay setting 
is zero).

External Pulse Signal from Machine/Sensor 

vs. Stroboscope Flash

Down Edge