background image

■ Measuring procedure

Install the stand body on a level solid table.

During measurement, operate the lever at a constant speed as smoothly as possible. If the lever operation speed is rapidly 

changed (to a high or low speed), accurate measurement cannot be performed.

During measurement and height adjustment, use thorough caution not to apply impact to the force gauge measuring axis.

The measuring stand can be horizontally installed.

To apply large load, fasten the stand body.

For details on the force gauge, refer to the force gauge operation manual for each model.

■ Stopper adjustment procedure

To  determine  the  force  gauge  lower  end  position  during  measurement, 

adjust the stopper position according to the procedure below.

Loosen the lock nut at the bottom of the slide assembly.

1) As the adjusting bolt is screwed into the slide assembly, the stopper 

position is raised.

2)  After  positioning  the  stopper,  tighten  the  lock  nut  to  complete  the 

stopper adjustment procedure.

■ Force gauge mounting procedure

1) Remove the mounting bolts by using the hexagon wrench (accessory), and
 remove the force gauge mounting plate.
2) Mount the force gauge to the force gauge mounting plate as shown below.

3) Mount the force gauge to the stand body by using the removed bolts.

FGC/FGN series

Cross-recessed 
pan head screw, 
M4 x 8 (4 pcs) 
(accessory)

FGX series

Cross-recessed 
pan head screw, 
M5 x 6 (3 pcs) 
(accessory)

Force gauge

Mounting plate

Mounting plate

Mounting plate

Slide assembly

Adjusting range

Lock nut

Adjusting bolt

Force gauge

:1 Terada Kotari, Nagaokakyo-city, Kyoto, 617-0833 Japan

 

Phone: 81-75-958-36