Thank you for purchasing this Clifton product. To get the best performance from the 
equipment, and for your personal safety, please read these instructions carefully before use. 

Clifton hotplates and stirrers are designed to operate under the following conditions:- 

 For indoor use only

❖ 

Use in a well ventilated area

❖ 

Ambient temperature range +5°C to +40°C

❖ 

Altitude to 2000m

❖ 

Relative humidity not exceeding 80%

❖ 

Mains supply fluctuations not exceeding 10% of nominal

❖ 

Overvoltage category II IEC60364-4-443

❖ 

Pollution degree 2 IEC664

❖ 

Use with a minimum distance all round of 200mm from walls or other items

If the equipment is not used in the manner described in this manual the protection provided by the 
equipment may be impaired.

Electrical Installation

THIS EQUIPMENT MUST BE EARTHED

Before connection please ensure that the line supply corresponds to that stated on the 
rating label.

Power requirements:

Model 

Wattage  

Model 

Wattage

 

CH-1E 

550W 

MSH-1 

750W

CHS-1E 

550W 

MSH-1D  750W

CHS-2E 

1200W 

MSH-3 

3 x 700W

There is an IEC socket at the rear of the instrument for connection to the mains supply, (see figure 2). 
The unit is supplied with two mains leads fitted with  IEC plugs for connection to the instrument.  
One has a U.K. 3 pin plug and the other has a 2 pin “Shuko” plug for connection to the mains.  
Choose the lead appropriate for your electrical installation and discard the other.

Should neither lead be suitable, take the lead with the U.K. plug and replace the plug with a suitable 
alternative. See the enclosed instruction sheet for advice on how to carry out this procedure.
Should the mains lead require replacement a cable of 1mm

2

 of harmonised code H05W-F connected 

to an IEC 320 plug should be used. 

IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN

The mains lead should be connected to the instrument BEFORE connection to the mains supply.
NOTE that when connected to the mains supply a green neon light on the front of the unit 
illuminates, (see figure 1).

The Clifton Range - 3 Year Warranty

3

High Quality Products at Affordable Prices

Содержание Clifton Series

Страница 1: ...rs THE CliftonRANGE Round and Rectangular Boiling Baths with Energy Regulator 1000ER and 1100ER Series PDF created with pdfFactory Pro trial version www software partners co uk MSH 1 MSH 1D MSH 3 CH 1...

Страница 2: ...ht Heater on light Stirrer on light Mains on light Hot warning light IEC socket External probe socket Retort rod fitting Digital display Stirrer control Heater control Digital models Figure 3 Front vi...

Страница 3: ...odel Wattage CH 1E 550W MSH 1 750W CHS 1E 550W MSH 1D 750W CHS 2E 1200W MSH 3 3 x 700W There is an IEC socket at the rear of the instrument for connection to the mains supply see figure 2 The unit is...

Страница 4: ...se the knob Repeat to turn off the heating function When switched on the display will show actual plate temperature at the bottom of the screen and the set temperature directly above The heating icon...

Страница 5: ...equately 2 Keep the unit out of strong drafts 3 Allow the temperature adequate time to stabilise Safety Advice Do not use the instrument to heat flammable liquids Never lift or carry the instrument un...

Страница 6: ...at glass bottles on a hotplate Ensure that the heat is built up slowly to avoid localised overheating Ensure the glassware is completely free from scratches or other defects Place the hotplate in a tr...

Страница 7: ...under normal laboratory conditions for a period of three 3 years In the event of a justified claim Nickel Electro will replace any defective component free of charge This warranty does NOT apply if 1...

Страница 8: ...of laboratory medical and clinical equipment Oldmixon Crescent Weston Super Mare North Somerset BS24 9BL United Kingdom Tel 44 0 1934 626691 Fax 44 0 1934 630300 Email clifton nickel electro co uk www...

Отзывы: