Thank you for purchasing this Clifton product. To get the best performance from the
equipment, and for your personal safety, please read these instructions carefully before use.
Clifton hotplates and stirrers are designed to operate under the following conditions:-
❖
For indoor use only
❖
Use in a well ventilated area
❖
Ambient temperature range +5°C to +40°C
❖
Altitude to 2000m
❖
Relative humidity not exceeding 80%
❖
Mains supply fluctuations not exceeding 10% of nominal
❖
Overvoltage category II IEC60364-4-443
❖
Pollution degree 2 IEC664
❖
Use with a minimum distance all round of 200mm from walls or other items
If the equipment is not used in the manner described in this manual the protection provided by the
equipment may be impaired.
Electrical Installation
THIS EQUIPMENT MUST BE EARTHED
Before connection please ensure that the line supply corresponds to that stated on the
rating label.
Power requirements:
Model
Wattage
Model
Wattage
CH-1E
550W
MSH-1
750W
CHS-1E
550W
MSH-1D 750W
CHS-2E
1200W
MSH-3
3 x 700W
There is an IEC socket at the rear of the instrument for connection to the mains supply, (see figure 2).
The unit is supplied with two mains leads fitted with IEC plugs for connection to the instrument.
One has a U.K. 3 pin plug and the other has a 2 pin “Shuko” plug for connection to the mains.
Choose the lead appropriate for your electrical installation and discard the other.
Should neither lead be suitable, take the lead with the U.K. plug and replace the plug with a suitable
alternative. See the enclosed instruction sheet for advice on how to carry out this procedure.
Should the mains lead require replacement a cable of 1mm
2
of harmonised code H05W-F connected
to an IEC 320 plug should be used.
IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN
The mains lead should be connected to the instrument BEFORE connection to the mains supply.
NOTE that when connected to the mains supply a green neon light on the front of the unit
illuminates, (see figure 1).