SCC-RLY01 Relay Module User Guide

4

ni.com

Figure 1.  

SCC-RLY01 Circuit Diagram

Caution

When connecting signals >60 VDC to the SCC-RLY01 module in the SCC-68, 

you must use the high-voltage backshell. Figure 2 shows the high-voltage backshell.

Figure 2.  

High-Voltage Backshell

For information about how to configure the SCC-RLY01 module with 
NI-DAQmx, refer to the 

SCC Quick Start Guide

.

Specifications

These ratings are typical at 25 °C unless otherwise stated.

Electrical

Contact type ............................................SPDT (Form C), nonlatching

Nominal switching capacity

SCC-68 ............................................5 A at 250 VAC

5 A at 30 VDC

SC-2345...........................................5 A at 30 VAC

5 A at 30 VDC

SCC-RLY01

E/M Series

NO

NC

COM

+5 V

P0.(

)

D GND

1

2

3

390 

, 1/4 W

Содержание SCC-RLY01

Страница 1: ...nt a range of values associated with a bit or signal name for example P0 3 0 The symbol leads you through nested menu items and dialog box options to a final action The sequence File Page Setup Option...

Страница 2: ...ries programs subprograms subroutines device names functions operations variables filenames and extensions SC 2345 SC 2345 refers to both the SC 2345 connector block and SC 2345 with configurable conn...

Страница 3: ...manuals Installing the Module Caution Refer to the Read Me First Safety and Radio Frequency Interference document before removing equipment covers or connecting disconnecting any signal wires Plug the...

Страница 4: ...kshell Figure 2 High Voltage Backshell For information about how to configure the SCC RLY01 module with NI DAQmx refer to the SCC Quick Start Guide Specifications These ratings are typical at 25 C unl...

Страница 5: ...e Signal bandwidth DC to 400 Hz Contact resistance 30 m max Switching time Operate time NC to NO 5 ms 10 ms max Release time NO to NC 4 ms 5 ms max 1 Maximum operating speed 30 cps at rated load Conta...

Страница 6: ...ion to signals in Measurement Category II III or IV Do not connect to MAINS When used with an SCC 68 Channel to earth inputs 300 VDC Measurement Category II1 Caution Do not use for connection to signa...

Страница 7: ...ts Minimum Immunity EN 55011 Emissions Group 1 Class A CE C Tick ICES and FCC Part 15 Emissions Class A Note For EMC compliance operate this device according to product documentation CE Compliance Thi...

Страница 8: ...on the bottom of the module Figure 4 SCC Module Bottom View Table 1 lists the signal connection corresponding to each pin GND is the reference for the 5 V supply 1 Pin 1 2 Pin 2 3 PWB Key 4 Pin 19 5 P...

Страница 9: ...any names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the appropriate location Help Patents in your softwar...

Отзывы: