NI PXIe-4154 Calibration Procedure

|

© National Instruments Corporation

|

25

Adjusting Current Measurement

To adjust the current measurement of a calibrated NI PXIe-4154, compare a range of current test 
points measured with an external DMM as a voltmeter in conjunction with a precision shunt to 
the measured current reported by the NI PXIe-4154. Refer to Figure 3 or Figure 4 to adjust 
current measurement while sourcing. Refer to Figure 5 or Figure 6 to adjust current 
measurement while sinking.

Refer to Table 17 for the current test points that you must request and measure with an external 
DMM and the NI PXIe-4154 to adjust a given range. For example, the 30 mA on Channel 0 
requires the adjustment application to separately request 27 mA, 0 mA, and -27 mA outputs from 
the NI PXIe-4154.

For positive currents, the voltage level should be set to the maximum value. Take measurements 
with the DMM/precision shunt and the NI PXIe-4154 at each test point. To calculate the current 
value using the DMM/precision shunt, divide the voltage measured with the DMM by the 
characterized value of the precision shunt.

For negative current adjustment points, an external current source is required to provide the 
necessary current. The voltage level for the NI PXIe-4154 should be set to 0 V. Refer to Table 1 
for external current source equipment recommendations and requirements.

Note

After requesting a new voltage, wait 3 ms to take a measurement with the 

external DMM and NI PXIe-4154 to ensure the system has adequate time to settle 
while minimizing self-heating effects of the internal measurement circuitry. 

Note

When measuring a 0 A test point, remove any connections to the front panel 

I/O connector to ensure no current is flowing through the output.

After test points are measured for a range, call the niDCPower Cal Adjust Current Measurement 
VI or the 

niDCPower_CalAdjustCurrentMeasurement

 function to calculate updated 

calibration coefficients. Refer to Table 18 for specific adjustment applications for this VI. Refer 
to the 

NI DC Power Supplies and SMUs Help 

for more information about programming 

references common to many VIs within the NI-DCPower API.

Table 17.  

NI PXIe-4154 Current Measurement Adjustment Points

Channel

I Range

Required Adjustment Points

Precision Shunt Value

0

30 mA

-27 mA

0 mA

27 mA

100 

0

3 A

-2 A

0 A

2 A

100 m

1

1.5 A

0 A

0.75 A

1.35 A

100 m

Содержание PXIe-4154

Страница 1: ...gulation Verification 17 Remote Sense Accuracy Verification 19 Transient Response Verification Channel 0 Only 20 Adjustment 23 Considerations 24 Adjusting Current Measurement 25 Adjusting Voltage Outp...

Страница 2: ...ice number for example 4154 and select the specifications document for your device Use this method any time the Start menu is referenced in this document Documentation Requirements For information abo...

Страница 3: ...cient Minimum power rating 0 5 W External Current Source Fluke 5500A or Fluke 5520A Current Measurement Current output of at least 3 A Less than 8 A RMS Normal Mode Current Noise Programmable Electron...

Страница 4: ...t least 30 minutes after the NI DCPower driver is loaded Unless manually disabled the NI DCPower driver automatically loads with the operating system and enables the device Allow the recommended warm...

Страница 5: ...wer Supplies and SMUs Getting Started Guide for information about how to install the software and hardware and how to configure the device in MAX Verification This section provides information for ver...

Страница 6: ...when measuring current output using a 0 1 shunt and voltage output for 0 V level Refer to Figure 2 for a connection diagram Figure 2 Voltage Verification Connection Diagram Note Use separate twisted...

Страница 7: ...00 m precision shunt Refer to Figure 4 for a connection diagram with a 100 precision shunt Figure 3 Current Verification Connection Diagram with 100 m Precision Shunt Table 2 NI PXIe 4154 Voltage Outp...

Страница 8: ...7 5 mA and 0 1 mA outputs from the NI PXIe 4154 while setting the voltage level to its maximum value Take measurements using the DMM as a voltmeter at each point dividing by the characterized value of...

Страница 9: ...t point using the DMM Compare these values to the voltage measurements taken at each test point using the NI PXIe 4154 Note After requesting a new voltage wait one second to take a measurement with an...

Страница 10: ...4 in order to complete verification To verify positive currents use the NI PXIe 4154 channel as a source For example in the 1 5 A range on Channel 1 program the NI PXIe 4154 to output 375 mA 750 mA 1...

Страница 11: ...nt is always zero Note After requesting a new current wait one second to take a measurement with an external DMM and NI PXIe 4154 to ensure the system has adequate time to settle The verification limi...

Страница 12: ...e external DMM or current source Refer to Figure 1 for a connection diagram Table 5 NI PXIe 4154 Current Measurement Verification Points Channel Range Precision Shunt Value Test Point Measured Current...

Страница 13: ...m 0 04 0 7 mA 1 5 A 0 065 0 7 mA 2 25 A 0 1025 0 7 mA 3 A 0 14 0 7 mA 1 1 5 A 100 m 0 A 0 12 0 35 mA 375 mA 0 12 0 35 mA 750 mA 0 17 0 35 mA 1 125 A 0 245 0 35 mA 1 5 A 0 32 0 35 mA Table 5 NI PXIe 41...

Страница 14: ...tronic load Refer to Figure 1 for a connection diagram Use short wires to jumper Sense to Output and Sense to Output Refer to Table 6 for a list of output resistance test points that you must request...

Страница 15: ...oltage measured across the 100 m precision shunt by the characterized value of the 100 m precision shunt 4 Increase the electronic load to 1 A constant current 5 Record the DMM voltage at the Output a...

Страница 16: ...oints Channel V Level I Limit Test Point Measured Voltages Measured Currents Limits of Accuracy Calculated Resistance V 1 V 2 I 1 I 2 Lower Limit Upper Limit As Found As Left 0 4 V 3 A 40 m 0 04312 0...

Страница 17: ...n diagram Note For Channel 0 of the NI PXIe 4154 ensure that output resistance is set to 0 during voltage load regulation verification Complete the following steps to verify voltage load regulation 1...

Страница 18: ...for a list of resistance values and measurements to complete verification Refer to Figure 7 for a connection diagram Note For each test the units for all current measurements and calculations should...

Страница 19: ...it Remote Sense Accuracy Verification To verify the remote sense accuracy of an NI PXIe 4154 use a test circuit of three resistors that simulate voltage drop between the device and a load Channels 0 a...

Страница 20: ...able electronic load with dynamic current capability Transient response validation should be performed only after successfully validating the voltage and current measurement accuracy and the load regu...

Страница 21: ...ransient response dip as displayed in Figure 10 The transient response recovery time is measured from the time the dip begins until the time at which the waveform crosses above the final voltage minus...

Страница 22: ...Fundamentals section of the NI DC Power Supplies and SMUs Help Figure 10 Example Transient Response Dip and Transient Response Recovery Time Waveform Note Measurement noise can be improved by software...

Страница 23: ...ts the following NI PXIe 4154 specifications Voltage programming accuracy Current programming accuracy Voltage measurement accuracy Current measurement accuracy Output resistance accuracy The adjustme...

Страница 24: ...e refer to the example programs installed with NI DCPower 3 Call niDCPower Cal Adjust function s When the measurements required for adjustment of a range are complete call one of the niDCPower Cal Adj...

Страница 25: ...the precision shunt For negative current adjustment points an external current source is required to provide the necessary current The voltage level for the NI PXIe 4154 should be set to 0 V Refer to...

Страница 26: ...ke measurements using the external DMM at each voltage test point Note Do not use the NI DCPower SFP to request test points for adjusting voltage outputs Note After requesting a new voltage wait one s...

Страница 27: ...shunt Refer to Figure 4 for a connection diagram with a 100 precision shunt Refer to Table 21 for a list of current test points that you must request and measure for each range For example the 30 mA r...

Страница 28: ...or the niDCPower_CalAdjustCurrentLimit function Figure 14 NI DCPower Cal Adjust Current Limit VI Block Diagram Table 21 NI PXIe 4154 Current Output Adjustment Points Channel I Range Required Adjustme...

Страница 29: ...st 0 1 V 4 V and 8 V outputs from the NI PXIe 4154 Take measurements using an external DMM and NI PXIe 4154 at each test point Note After requesting a new voltage wait one second to take a measurement...

Страница 30: ...he electronic load to an initial constant current of 100 mA Obtain V1 and I1 by measuring the voltage and calculating the current using the external DMMs and precision shunt Use the following steps to...

Страница 31: ...e 25 NI PXIe 4154 Output Resistance Adjustment Points Channel V Range V Level I Limit Required Adjustment Points 0 6 V 4 V 3 A 0 0 5 0 9 Table 26 NI DCPower Cal Adjust Output Resistance VI Parameter D...

Страница 32: ...export data NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS U S Government Customers The data contained in this...

Отзывы: