NI PXIe-4154 Calibration Procedure

The verification limits for voltage measurement depend on the actual voltage measured with an 
external DMM, and are expressed as a percentage of DMM reading plus offset voltage. Refer to 
Table 4 for more information.

Current Measurement Verification

To verify the current measurement of an NI PXIe-4154, measure a range of current test points 
with an external DMM and a precision shunt. Divide the voltage across the precision shunt by 
the actual resistance of the precision shunt. Compare these values to the values of the current 
measured at the same current test points using the NI PXIe-4154. Refer to Figure or Figure 
for the configuration to verify current measurements while sourcing. Refer to Figure 5 or 
Figure 6 for the configuration to verify current measurement while sinking.

Refer to Table 5 for a list of current test points for each range that you must request and measure 
with an external DMM/precision shunt and the NI PXIe-4154 in order to complete verification.

To verify positive currents, use the NI PXIe-4154 channel as a source. For example, in the 1.5 A 
range on Channel 1, program the NI PXIe-4154 to output 375 mA, 750 mA, 1.125 A, and 1.5 A 
while setting the voltage limit to its maximum value. Use an external DMM/precision shunt and 
the NI PXIe-4154 to take a current measurement at each test point. Compare the measured 
values against each other.

To verify negative currents, use an external current source to force Channel 0 of the 
NI PXIe-4154 to sink current. For example, in the 3 A range on Channel 0, program the 
NI PXIe-4154 channel for 0 V output and program the external current source to output 750 mA, 
1.5 A, 2.25 A, and 3 A. Channel 0 will be sinking current, so its measurements will be negative 
currents. Take a current measurement at each test point using both an external DMM/precision 
shunt and the NI PXIe-4154. Compare the measured values against each other.

Table 4.  

NI

PXIe-

4154 Voltage Measurement Verification Points

Channel

Range

Test 

Point

Measured Voltage

Test Limit

Reported Voltage

As 

Found

As 

Left

As 

Found

As 

Left

0

6 V

0 V

0.025% + 2 mV

1.5 V

3 V

4.5 V

6 V

1

8 V

0 V

0.05% + 2.5 mV

2 V

4 V

6 V

8 V

Содержание PXIe-4154

Страница 1: ...gulation Verification 17 Remote Sense Accuracy Verification 19 Transient Response Verification Channel 0 Only 20 Adjustment 23 Considerations 24 Adjusting Current Measurement 25 Adjusting Voltage Outp...

Страница 2: ...ice number for example 4154 and select the specifications document for your device Use this method any time the Start menu is referenced in this document Documentation Requirements For information abo...

Страница 3: ...cient Minimum power rating 0 5 W External Current Source Fluke 5500A or Fluke 5520A Current Measurement Current output of at least 3 A Less than 8 A RMS Normal Mode Current Noise Programmable Electron...

Страница 4: ...t least 30 minutes after the NI DCPower driver is loaded Unless manually disabled the NI DCPower driver automatically loads with the operating system and enables the device Allow the recommended warm...

Страница 5: ...wer Supplies and SMUs Getting Started Guide for information about how to install the software and hardware and how to configure the device in MAX Verification This section provides information for ver...

Страница 6: ...when measuring current output using a 0 1 shunt and voltage output for 0 V level Refer to Figure 2 for a connection diagram Figure 2 Voltage Verification Connection Diagram Note Use separate twisted...

Страница 7: ...00 m precision shunt Refer to Figure 4 for a connection diagram with a 100 precision shunt Figure 3 Current Verification Connection Diagram with 100 m Precision Shunt Table 2 NI PXIe 4154 Voltage Outp...

Страница 8: ...7 5 mA and 0 1 mA outputs from the NI PXIe 4154 while setting the voltage level to its maximum value Take measurements using the DMM as a voltmeter at each point dividing by the characterized value of...

Страница 9: ...t point using the DMM Compare these values to the voltage measurements taken at each test point using the NI PXIe 4154 Note After requesting a new voltage wait one second to take a measurement with an...

Страница 10: ...4 in order to complete verification To verify positive currents use the NI PXIe 4154 channel as a source For example in the 1 5 A range on Channel 1 program the NI PXIe 4154 to output 375 mA 750 mA 1...

Страница 11: ...nt is always zero Note After requesting a new current wait one second to take a measurement with an external DMM and NI PXIe 4154 to ensure the system has adequate time to settle The verification limi...

Страница 12: ...e external DMM or current source Refer to Figure 1 for a connection diagram Table 5 NI PXIe 4154 Current Measurement Verification Points Channel Range Precision Shunt Value Test Point Measured Current...

Страница 13: ...m 0 04 0 7 mA 1 5 A 0 065 0 7 mA 2 25 A 0 1025 0 7 mA 3 A 0 14 0 7 mA 1 1 5 A 100 m 0 A 0 12 0 35 mA 375 mA 0 12 0 35 mA 750 mA 0 17 0 35 mA 1 125 A 0 245 0 35 mA 1 5 A 0 32 0 35 mA Table 5 NI PXIe 41...

Страница 14: ...tronic load Refer to Figure 1 for a connection diagram Use short wires to jumper Sense to Output and Sense to Output Refer to Table 6 for a list of output resistance test points that you must request...

Страница 15: ...oltage measured across the 100 m precision shunt by the characterized value of the 100 m precision shunt 4 Increase the electronic load to 1 A constant current 5 Record the DMM voltage at the Output a...

Страница 16: ...oints Channel V Level I Limit Test Point Measured Voltages Measured Currents Limits of Accuracy Calculated Resistance V 1 V 2 I 1 I 2 Lower Limit Upper Limit As Found As Left 0 4 V 3 A 40 m 0 04312 0...

Страница 17: ...n diagram Note For Channel 0 of the NI PXIe 4154 ensure that output resistance is set to 0 during voltage load regulation verification Complete the following steps to verify voltage load regulation 1...

Страница 18: ...for a list of resistance values and measurements to complete verification Refer to Figure 7 for a connection diagram Note For each test the units for all current measurements and calculations should...

Страница 19: ...it Remote Sense Accuracy Verification To verify the remote sense accuracy of an NI PXIe 4154 use a test circuit of three resistors that simulate voltage drop between the device and a load Channels 0 a...

Страница 20: ...able electronic load with dynamic current capability Transient response validation should be performed only after successfully validating the voltage and current measurement accuracy and the load regu...

Страница 21: ...ransient response dip as displayed in Figure 10 The transient response recovery time is measured from the time the dip begins until the time at which the waveform crosses above the final voltage minus...

Страница 22: ...Fundamentals section of the NI DC Power Supplies and SMUs Help Figure 10 Example Transient Response Dip and Transient Response Recovery Time Waveform Note Measurement noise can be improved by software...

Страница 23: ...ts the following NI PXIe 4154 specifications Voltage programming accuracy Current programming accuracy Voltage measurement accuracy Current measurement accuracy Output resistance accuracy The adjustme...

Страница 24: ...e refer to the example programs installed with NI DCPower 3 Call niDCPower Cal Adjust function s When the measurements required for adjustment of a range are complete call one of the niDCPower Cal Adj...

Страница 25: ...the precision shunt For negative current adjustment points an external current source is required to provide the necessary current The voltage level for the NI PXIe 4154 should be set to 0 V Refer to...

Страница 26: ...ke measurements using the external DMM at each voltage test point Note Do not use the NI DCPower SFP to request test points for adjusting voltage outputs Note After requesting a new voltage wait one s...

Страница 27: ...shunt Refer to Figure 4 for a connection diagram with a 100 precision shunt Refer to Table 21 for a list of current test points that you must request and measure for each range For example the 30 mA r...

Страница 28: ...or the niDCPower_CalAdjustCurrentLimit function Figure 14 NI DCPower Cal Adjust Current Limit VI Block Diagram Table 21 NI PXIe 4154 Current Output Adjustment Points Channel I Range Required Adjustme...

Страница 29: ...st 0 1 V 4 V and 8 V outputs from the NI PXIe 4154 Take measurements using an external DMM and NI PXIe 4154 at each test point Note After requesting a new voltage wait one second to take a measurement...

Страница 30: ...he electronic load to an initial constant current of 100 mA Obtain V1 and I1 by measuring the voltage and calculating the current using the external DMMs and precision shunt Use the following steps to...

Страница 31: ...e 25 NI PXIe 4154 Output Resistance Adjustment Points Channel V Range V Level I Limit Required Adjustment Points 0 6 V 4 V 3 A 0 0 5 0 9 Table 26 NI DCPower Cal Adjust Output Resistance VI Parameter D...

Страница 32: ...export data NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS U S Government Customers The data contained in this...

Отзывы: