Overview

This document explains how to connect to the NI 9205.

Note

 Before you begin, complete the software and hardware installation

procedures in your chassis documentation.

Note

 The guidelines in this document are specific to the NI 9205. The

other components in the system might not meet the same safety ratings.

Refer to the documentation for each component in the system to

determine the safety and EMC ratings for the entire system.

Safety Guidelines

Operate the NI 9205 only as described in this document.

Caution

 This icon denotes a caution, which advises you to consult

documentation where this symbol is marked.

Caution

 Do not operate the NI 9205 in a manner not specified in this

document. Product misuse can result in a hazard. You can compromise the

safety protection built into the product if the product is damaged in any

way. If the product is damaged, return it to NI for repair.

© National Instruments

3

NI-9205 Getting Started

Содержание NI-9205

Страница 1: ...NI 9205 Getting Started 2022 07 06...

Страница 2: ...pecial Conditions for Marine Applications 8 Preparing the Environment 8 Connecting the NI 9205 9 NI 9205 Signal Descriptions 9 Signals 10 Connecting Grounded Differential Signals 10 Connecting Floatin...

Страница 3: ...system to determine the safety and EMC ratings for the entire system Safety Guidelines Operate the NI 9205 only as described in this document Caution This icon denotes a caution which advises you to c...

Страница 4: ...Voltages Channel to channel None Channel to earth ground Continuous 250 V RMS Measurement Category II Withstand up to 2 000 m 3 000 V RMS verified by a 5 s dielectric withstand test Measurement Catego...

Страница 5: ...ements include signal levels special equipment limited energy parts of equipment circuits powered by regulated low voltage sources and electronics Caution Do not connect the NI 9205 with DSUB to signa...

Страница 6: ...not accessible Safety Guidelines for Hazardous Locations The NI 9205 is suitable for use in Class I Division 2 Groups A B C D T4 hazardous locations Class I Zone 2 AEx nA IIC T4 Gc and Ex nA IIC T4 G...

Страница 7: ...T4 or Ex nL IIC T4 equipment Caution Transient protection shall be provided that is set at a level not exceeding 140 of the peak rated voltage value of 85 V at the supply terminals to the equipment Ca...

Страница 8: ...itions for Marine Applications Some products are Lloyd s Register LR Type Approved for marine shipboard applications To verify Lloyd s Register certification for a product visit ni com certification a...

Страница 9: ...22 AI23 AISENSE 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 AI0 AI1 AI3 AI5 AI7 AI17 AI19 AI21 AI23 DO0 AI2 AI4 AI6 AI16 AI18 AI20 AI22 COM AI...

Страница 10: ...tain more accurate measurements and less noise The following table shows the signal pairs that are valid for differential connection configurations with the NI 9205 Channel AI AI 0 AI0 AI8 1 AI1 AI9 2...

Страница 11: ...ecting Floating Differential Signals to the NI 9205 NI 9205 MUX ADC PGIA Vcm AI0 CH0 AI8 CH0 AI1 CH1 AI9 CH1 COM V1 V2 Connecting RSE Voltage Signals You can connect referenced single ended RSE signal...

Страница 12: ...G to 18 AWG copper conductor wire with 7 mm 0 28 in of insulation stripped from the end Flathead screwdriver with a 2 3 mm x 1 0 mm 0 09 in x 0 04 in blade included with the NI 9205 What to Do Complet...

Страница 13: ...VIEW FPGA Help NI 9205 Datasheet NI DAQmx Help LabVIEW Help Installs with the software Located at ni com manuals Worldwide Support and Services The NI website is your complete resource for technical s...

Страница 14: ...your product at ni com calibration NI corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 NI also has offices located around the world For telephone support in t...

Отзывы: