NI 5421/5441 Calibration Procedure
b.
Call
niFgen_SetAttributeViReal64
to set the digital filter interpolation factor
(NI-FGEN Digital Filter Interpolation Factor property) using the following
parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_DIGITAL_FILTER_INTERPOLATION_FACTOR
•
value
:
4
15. Call
niFgen_InitiateGeneration
(niFgen Initiate Generation VI) using the
following parameter:
•
vi
: The session handle returned from
niFgen_init
16. Allow the power meter to stabilize for 10 seconds.
17. Measure and record the power (
W
f
) of the positive output in Watts. Use the recorded power
at 50 kHz as reference power (
W
ref
).
18. Call
niFgen_AbortGeneration
(niFgen Abort Generation VI) to abort the current
generation using the following parameter:
•
vi
: The session handle returned from
niFgen_init
19. For iterations 2-7 in Table 8, using the recorded power values, calculate the deviation from
the reference (50 kHz) power using the following equation:
20. Compare the
Flatness (dB)
value for the
current amplifier path listed in Table 8.
21. To verify the flatness for the High-Gain Amplifier Path, repeat the process from step 1, but
in step 8, set the
Gain
to
3
.
22. Call
niFgen_close
(niFgen Close VI) to close the instrument driver session, to destroy
the instrument driver session and all of its properties, and to release any memory resources
that NI-FGEN uses. Use the following parameter:
•
vi
: The session handle returned from
niFgen_init
Verifying the Direct Analog Path Flatness
Complete the following steps to verify the direct analog path flatness of the NI 5421/5441,
complete the following steps:
1.
Connect the NI 5421/5441 CH 0 front panel connector to the power meter using the
required adapter.
2.
Call
niFgen_init
(niFgen Initialize VI) using the following parameters:
•
resourceName
: The name of the device that you want to verify. You can find this
name under Devices and Interfaces in MAX.
•
IDQuery
:
VI_TRUE
•
resetDevice
:
VI_TRUE
Flatness dB
10
W
f
W
ref
------------
log
=