26
|
ni.com
|
NI 5421/5441 Calibration Procedure
•
attributeID
:
NIFGEN_ATTR_ANALOG_PATH
•
value
:
NIFGEN_VAL_MAIN_ANALOG_PATH
8.
Call
niFgen_SetAttributeViReal64
to set the gain (NI-FGEN Gain property) using
the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_ARB_GAIN
•
value
:
1
9.
Call
niFgenSetAttributeViBoolean
(NI-FGEN Output Enabled property) to enable
the NI 5421/5441 output. Use the following parameters:
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_OUTPUT_ENABLED
•
value
:
VI_TRUE
•
vi
: The session handle returned from
niFgen_init
10. Call
niFgen_ConfigureSampleRate
(niFgen Set Sample Rate VI) using the following
parameters:
•
vi
: The session handle returned from
niFgen_init
•
Sample Rate
: 100 MS/s (
100000000
)
Repeat steps 11 through 20 for each iteration in Table 8, changing the
for each iteration.
Table 8.
NI 5421/5441 Setup for Main Analog Path Flatness Verification
Iteration
Frequency
Number of
Samples
Number of
Cycles
Published Specification
Low-Gain
Amplifier
High-Gain
Amplifier
1
50 kHz
2,000
1
REF
REF
2
100 kHz
1,000
1
-1.0 dB to
+0.5 dB
-1.2 dB to
+0.5 dB
3
1 MHz
1,000
10
-1.0 dB to
+0.5 dB
-1.2 dB to
+0.5 dB
4
5 MHz
1,000
50
-1.0 dB to
+0.5 dB
-1.2 dB to
+0.5 dB
5
10 MHz
1,000
100
-1.0 dB to
+0.5 dB
-1.2 dB to
+0.5 dB
6
15 MHz
1,000
150
-1.0 dB to
+0.5 dB
-1.2 dB to
+0.5 dB