This device is intended for use in the electromagnetic environment specified below. The
user of this device should make sure it is used in such an environment.
GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC IMMUNITY:
Electrostatic
Discharge
(ESD)IEC
61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
±8kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV for
input-output lines
±1 kV
differential mode
±2 kV for
common mode
<5% UT (>95% dip in UT)
for 0.5 cycle
40%UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
<5% UT (>95% dip in UT)
for 0.5 cycle
40%UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
±1 kV
differential mode
±2 kV for
common mode
±2 kV for power
supply lines
±1 kV for
input-output lines
±8kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
shouldbe that of a typical
home or hospital
environment.
Mains power quality should
be that of a typical home
orhospital environment.
Mains power quality
should be that of a typical
home or hospital
environment.
If the user of the device
requires continued
operation during power
mains interruptions, it is
recommended that the
device be powered from
an uninterruptible power
supplyor a battery.
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE LEVEL
ELECTROMAGNETIC
ENVIRONMENTGUIDANCE