6.1 Error messag
6-4
WF1946B
continued from previous page
Error message
Cause or corrective measures
STORE/RECALL MEMORY
LOST
Setting storage memory contents destroyed and settings cannot be
recalled.
Contact dealer.
WARNING 001
Combined frequency and duty set the pulse width to not more than
25 ns and the pulse may be lost.
WARNING 002
Combined frequency and duty set the pulse width to 25 and 100 ns
and pulse width may be unstable (large jitter component).
WARNING 003
Since high frequency, burst oscillation mark and space may be
unstable.
WARNING 004
Low level setting changed due to high level setting change, or
conversely, high level setting changed due to low level setting
change.
WARNING 005
Changed to simple standard units (Hz, s, Vp-p, V).
WARNING 006
Since combined period and pulse width exceed the permissible
duty setting range, pulse width was changed in order to enter the
permissible duty range.
WARNING 007
Changing the frequency relationship exceeded the permissible
frequency setting range for the other channel. Therefore, the
frequency relationship of both channels was changed to within this
range.
WARNING 008
Because of channel mode change, the oscillation mode was
changed to normal.
Содержание WF1946B
Страница 2: ......
Страница 3: ...DA00012506 002 WF1946B MULTIFUNCTION SYNTHESIZER Instruction Manual ...
Страница 4: ......
Страница 8: ......
Страница 28: ......
Страница 66: ......
Страница 107: ...4 2 Sweep WF1946B 4 41 ...
Страница 164: ......
Страница 180: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 203: ......
Страница 204: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN ...