4.2 Sweep
WF1945B
4-21
• FUNCTION
determines the sweep type.
: For example,
provides step-type change of output (e.g., frequency) at the sweep time halfway
point. LIN/LOG provides linear or logarithmic variation of the output with respect to the time axis.
i
i
i
i
SIN ( )
LIN
LOG
LIN
LOG
LIN
LOG
• Internal sweep trigger source
: The trigger signal is generated at the following rate, then sweep is conducted.
However, the trigger interval is 100 ms even if set to less.
INT
100.0ms
F-SWP:
TYPE
SOURCE
MODE
4
• START-STATE and STOP-STATE
: Respectively set the output to start and stop values.
Since the sweep auxiliary output becomes the respective start
and stop states, the full scale of the recorder can be adjusted
and the state of external equipment can be checked. The
Start-State is the same as pressing the
STOP
key (during single sweep).
Relation ship between sweep values and sweep auxiliary output
"
4.2 Sweep (
■
Sweep value and Z-MARKER/SYNC/
X- DRIVE outputs)", cf.
•
Sweep type set to duty
: Output waveform fixed to squarewave (variable duty) and waveform (Function) cannot be
selected.
During sweep, multiple pulses may be produced within 1 period output as indicated in the
figure.
• To generate the trigger signal via remote control (GPIB or USB),
set the TRG command from the
remote control interface.
For details of the remote control command,
“Remote Control Instruction Manual”, cf
.
Trigger rate
Содержание WF1945B
Страница 2: ......
Страница 3: ...DA00012505 002 WF1945B MULTIFUNCTION SYNTHESIZER Instruction Manual...
Страница 4: ......
Страница 8: ......
Страница 20: ......
Страница 30: ......
Страница 62: ......
Страница 100: ...4 2 Sweep 4 38 WF1945B Sweep value and Z MARKER SYNC X DRIVE outputs is approx 100 to 125 s...
Страница 101: ...4 2 Sweep WF1945B 4 39...
Страница 128: ......
Страница 158: ......
Страница 182: ......
Страница 184: ......
Страница 186: ......
Страница 187: ......
Страница 188: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...