4.2 Sweep
4-20
WF1945B
Other operations:
• Sweep stop
: Press the
STOP
key. When sweep is stopped at the stop value at single sweep
end or during sweep, press the
STOP
key to set the start value.
• Sweep pause
: Press the
PAUSE
key. Again press the
PAUSE
key to resume sweep.
• Single sweep operation examples
• Dead time and trigger delay
: Triggers input within 100 ms of completion of the last sweep are ignored.
The trigger delay for the first sweep after power-on or a parameter change is 40 ms.
Otherwise, it is 2 ms.
Power-on or a sweep parameter
setting change
Stop value
Dead time = 100 ms
Start value
Trigger
Trigger
Triggers during dead
time are ignored.
40ms
2ms
• External sweep start
: Set the sweep trigger source to EXT and select rising or falling. Supply an
external signal to the front panel TRIG/SWEEP IN connector. However, note that
retrigger cannot be applied for 100 ms after sweep start.
During a sweep, the sweep restarts if a trigger is input.
EXT
F-SWP:
TYPE
SOURCE
MODE
4
•
External sweep pause
: Supply a low level signal to the rear panel SWEEP PAUSE IN connector. Note
that sweep restarts at high level or open input.
Содержание WF1945B
Страница 2: ......
Страница 3: ...DA00012505 002 WF1945B MULTIFUNCTION SYNTHESIZER Instruction Manual...
Страница 4: ......
Страница 8: ......
Страница 20: ......
Страница 30: ......
Страница 62: ......
Страница 100: ...4 2 Sweep 4 38 WF1945B Sweep value and Z MARKER SYNC X DRIVE outputs is approx 100 to 125 s...
Страница 101: ...4 2 Sweep WF1945B 4 39...
Страница 128: ......
Страница 158: ......
Страница 182: ......
Страница 184: ......
Страница 186: ......
Страница 187: ......
Страница 188: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...