PPA45xx Quick User Guide
Page 113
Connecting up the PPA for “internal” measurements.
The breakout box is connected to a mains supply. A load is connected to the
breakout box to produce a current for the PPA to monitor. The PPA is used to
monitor the voltage and current on the connections of the breakout box. The same
voltage and current levels are applied to the 3 phase inputs of the PPA. Therefore
the display should indicate the nominally same values for all 3 phases. For the tests
in this document a 230V 50Hz mains supply was used, the load used was sinking a
current of 3.00A.
To make the connections shown in the diagram you will need the following
accessories that are not supplied with the PPA.
1 x breakout box.
1 x Load that will be connected to the breakout box.
All other connections can be made using the 4m leads supplied with the PPA.
Содержание PPA15 Series
Страница 1: ...PPA45xx series START UP GUIDE Firmware v2_169 06th November 2017...
Страница 81: ...PPA45xx Quick User Guide Page 80 Step 4 Click the Set up PPA button Step 5 Click the Start button...
Страница 108: ...PPA45xx Quick User Guide Page 107 Connection diagram for the functionality checks of the External BNC inputs...
Страница 115: ...PPA45xx Quick User Guide Page 114 Connection diagram for the functionality checks of the Internal 4mm sockets inputs...
Страница 117: ...PPA45xx Quick User Guide Page 116 True RMS Voltmeter Mode Power Analyzer Mode...