2D/3D OPT E.L. CALIBRATION MANUAL
Remove any object between source and sensor and
press “Start”
Press the button when prompted.
Check that the entire length of the X-ray beam is cen-
tred inside the red lines (compare with the following
images), in particular concerning rotation, horizontal
alignment and vertical alignment.
If adjustments are necessary, open the generator co-
ver and operate on the collimator as follows:
Upper part
Lower part