26
4.3.6 Sensor Parameters
• In the first column of Configuration/Device Setup menu (
Figure 4.11
), click on No.
1
to view and modify the Sensor Parameters Menu for Temperature.
Figure 4.12 Sensor Parameters Menu
Below are some definitions of terms used in the Sensor Parameters Menu.
Device Name:
shows on the Configuration and the Read Sensor page.
Display Format:
Decimal
- the leading zero is eliminated.
Raw
- the leading zero is there (do not change, for future use)
Remote Display Format:
Remote display format is for Terminal Server continuous mode.
H_00.00%
in humidity setting, displays
H_
and
00.00%
is the humidity value displayed.
Example: if the humidity is
30.00
, then
H_30.00%
will be seen in the remote display. If no
format is specified (blank), there is no reading sent out. If temperature is
76.25
and
00.0F
is used in temperature setting, the remote site will show
76.2F
.
This format setup is originally based on iLD Remote Display, which has four and six
LEDs. For six LEDs,
T_00.00F
can be set, and for four LEDs,
00.0F
can be used.
Remote End Char:
The default value is 0D (Hex representation of <CR>). This means
that the iServer sends <CR> after each temperature, humidity, and dewpoint value. This
will be done either in Continuous or Command mode.
This is how the data will appear on the host:
T_75.70F
H_37.90%
D_44.90F
If the end character is 20 (Hex representation of space), the
data will then appear as:
T_75.70F H_37.90% D_44.90F
If nothing is set for the “Remote End Char” field, the iServer will
then forward the data to the LAN with no characters followed.
Offset:
SInce the sensing probe is solid state electronics, there
is no need for calibration. If it’s determined that the readings
are slightly off, the user can manually assign numerical values
to adjust the readings for temperature, humidity and dewpoint.
For temperature offset, the unit must be in degree C. The
Offset can be changed while recording is on.
SENSOR PARAMETERS
Device No. 1
Device Name:
Main Menu
http://128.100.101.254
Sensor Parameter
Update
Reset
Temperature
Display Format:
decimal
Remote Display Format:
T_00.00F
Remote End Char (HEX) 0X:
0D
Offset:
( C or %)
002.0
Cancel
T_75.70F
H_37.90%
D_44.90F
T_75.70F
H_37.90%
D_44.90F
T_75.70F
H_38.00%
D_44.90F
T_75.70F
H_37.90%
D_44.90F
Tera Term - 206.29.25.27 VT
File Edit Setup Control Window Help
Figure 4.13
Remote End Char
You may
type any
ASCII
characters in the
first three boxes,
but you must
delete the leading
spaces.