31
READING ADJUST SUBMENU:
Press
d
10)
Display flashes 1
st
digit of previous reading adjust value.
Press
b
&
c
11)
Press
b
and
c
buttons to enter a new Reading Adjust value
(-1999 to 9999).
Press
d
12)
Display shows
STRD
stored message momentarily and then
advances to
SP.DV
Setpoint Deviation Menu.
Reading Offset Adjust
allows the user to fine tune a minor error of the transducer,
however some applications may require a large offset adjust.
(Displayed Process Value = Measured Process Value ± R.ADJ).
Reading Adjust is adjustable between -1999 to 9999. For Temperature reading only, not
Process.
SETPOINT DEVIATION ENABLE/DISABLE SUBMENU:
Press
d
13)
Display advances to Setpoint Deviation
ENBL
Enable or
DSBL
Disable Submenu and flashes the previous selection.
Press
b
14)
Scroll through the available selections:
ENBL
or
DSBL
.
Press
d
15)
Display shows
STRD
stored message momentarily and then
advances to the
CAL1
Menu.
Setpoint Deviation
Submenu, if “enabled”, allows changes to Setpoint 1 to be made
automatically to Setpoint 2. This mode is very helpful if the Process Value changes often.
In Setpoint Deviation Mode, set SP2 a certain number of degrees or counts away from
SP1 - this relation remains fixed when SP1 is changed. For instance: Setting SP1=200
and SP2=20 and enabling
SP.DV
means that the absolute value of SP2=20. Moving SP1
to 300, the absolute value of SP2 becomes 320.
THERMOCOUPLE FIELD CALIBRATION SUBMENU:
RTD and Process are perfectly calibrated. This section is applicable to
Thermocouple (TC) calibration
only
.
Be sure that the TC being used to calibrate the meter is of the type
selected in the TC submenu. Place the TC in an ice-bath (or other
0°C / 32°F
environment). In ambient temperature conditions: connect the
TC to the meter, apply power to the meter.
CAUTION:
Do not proceed with TC calibration unless the above conditions
have been in effect for at least one hour.
Press
a
16)
Display shows
CAL1
.
Press
d
17)
Display shows flashing
0000
.
Press
a
18)
Display will still show flashing
0000
.
Press
d
19)
Display shows
OUT1
(meaning Calibration is complete)
*
If you accidently engage the flashing
0000
(CAL° alert) simply re-press
the last button you pressed, to avoid unintentionally mis-calibrating your meter.
°°
CAUTION:
Do not perform the following steps until you fully
understand this entire section.
°°