6
Voltage Jumper Pin Settings
Figure 3-1 shows the location of solder jumpers W1 through W3.
Figure 3-1. PC Board Jumper Locations
S3
P3
P1
W1
W2
W3
2
1
2
3
1
TRANSFORMER
SPAN
DISPLA
Y
ZERO
(COMPONENT SIDE)
P2
2
3
1
AC VOLTAGE
INSTALL
REMOVE
115 Vac
W1, W3
W2
230 Vac
W2
W1, W3
23
TABLE 4-1 (Cont’d)
†
The best MID-RANGE accuracy is obtained when the
meter is calibrated to the exact values shown in column 3,
even though a small error may occur at 32°F (0°C).
*
Look at Table 1-1, for complete models (preface section).
MODEL
T/C CALIBRATOR METER
POT TO
OPTION
DISPLAY
ADJUST
DISPLAY
*-TF1
32°F
31°F †
Zero
716°F
716°F
Span
*-TF2
32.0°F
32.0°F
Zero
185.0°F
185.0°F
Span
*-TC1
0°C
0°C
Zero
380°C
380°C
Span
*-TC2
0.0°C
0.0°C
Zero
185.0°C
185.0°C
Span
*-EF1
32°F
31°F †
Zero
1769°F
1769°F
Span
*-EF2
32.0°F
33.0°F †
Zero
180.0°F
180.0°F
Span
*-EC1
0°C
0°C
Zero
965°C
965°C
Span
*-EC2
0°C
0.5°C †
Zero
185.0°C
185.0°C
Span
6
Voltage Jumper Pin Settings
Figure 3-1 shows the location of solder jumpers W1 through W3.
Figure 3-1. PC Board Jumper Locations
S3
P3
P1
W1
W2
W3
2
1
2
3
1
TRANSFORMER
SPAN
DISPLA
Y
ZERO
(COMPONENT SIDE)
P2
2
3
1
AC VOLTAGE
INSTALL
REMOVE
115 Vac
W1, W3
W2
230 Vac
W2
W1, W3
23
TABLE 4-1 (Cont’d)
†
The best MID-RANGE accuracy is obtained when the
meter is calibrated to the exact values shown in column 3,
even though a small error may occur at 32°F (0°C).
*
Look at Table 1-1, for complete models (preface section).
MODEL
T/C CALIBRATOR METER
POT TO
OPTION
DISPLAY
ADJUST
DISPLAY
*-TF1
32°F
31°F †
Zero
716°F
716°F
Span
*-TF2
32.0°F
32.0°F
Zero
185.0°F
185.0°F
Span
*-TC1
0°C
0°C
Zero
380°C
380°C
Span
*-TC2
0.0°C
0.0°C
Zero
185.0°C
185.0°C
Span
*-EF1
32°F
31°F †
Zero
1769°F
1769°F
Span
*-EF2
32.0°F
33.0°F †
Zero
180.0°F
180.0°F
Span
*-EC1
0°C
0°C
Zero
965°C
965°C
Span
*-EC2
0°C
0.5°C †
Zero
185.0°C
185.0°C
Span