11
7 Filter Element Replacement
Replace the filter element with a new one, if it is dirty or wet.
(1) Turn the retainer and short probe as a whole counterclockwise to remove
them from the filter case.
(2) Remove the O-ring from the filter case.
(3) Replace the filter element with a new one.
(4) Reinstall the O-ring back onto the case.
(5) Reinstall the retainer and short probe to the filter case.
Filter case
Filter element (FE-112)
O-ring
Retainer
Rotate to remove
Short probe
Press the
▲
or
▼
button few times during normal operation, the
following values will be displayed on the right bottom corner of
the LCD. They are just the ex-factory parameters and not for
customer use.
NOTE
10
6 Data Logging
The detector can log up to 22 hours of the measurement data. The log data
can also be transferred to a pc by using data logger kit (sold separately, see 10.
Consumable and Optional Parts on page 13).
(1) During normal operation, press the
▼
button.
“LoG” appears on the LCD. The log starting time
entry screen appears.
(2) Press the logger button. One of the four digits
flashes at one time. Enter the value for the
flashing digit with the
▲
and
▼
buttons. Repeat
this step four times to enter the four digits (log
starting time).
For example, when you want to record that the
starting time is 1:20 pm, enter “13.20”.
(3) Press and hold the ZERO button to start logging.
The logging starts.
“MNT1” appears while logging is in progress.
(4) To end the logging, press the BZ.STOP button.
“MNT1” disappears and the logging ends.
•
Once data logging starts, the peak concentration value
every 10 seconds will be logged.
•
The oldest events will be deleted to make room for
the newest when the unit runs out of room for new ones.
•
Data logging automatically stops when the detector is turned
off.
•
Logging date cannot be entered.
•
Refer to the XPS7L data logger software’s instruction
manual (separate document) for the procedure to transfer
log data to a pc.
NOTE
11
7 Filter Element Replacement
Replace the filter element with a new one, if it is dirty or wet.
(1) Turn the retainer and short probe as a whole counterclockwise to remove
them from the filter case.
(2) Remove the O-ring from the filter case.
(3) Replace the filter element with a new one.
(4) Reinstall the O-ring back onto the case.
(5) Reinstall the retainer and short probe to the filter case.
Filter case
Filter element (FE-112)
O-ring
Retainer
Rotate to remove
Short probe
Press the
▲
or
▼
button few times during normal operation, the
following values will be displayed on the right bottom corner of
the LCD. They are just the ex-factory parameters and not for
customer use.
NOTE
10
6 Data Logging
The detector can log up to 22 hours of the measurement data. The log data
can also be transferred to a pc by using data logger kit (sold separately, see 10.
Consumable and Optional Parts on page 13).
(1) During normal operation, press the
▼
button.
“LoG” appears on the LCD. The log starting time
entry screen appears.
(2) Press the logger button. One of the four digits
flashes at one time. Enter the value for the
flashing digit with the
▲
and
▼
buttons. Repeat
this step four times to enter the four digits (log
starting time).
For example, when you want to record that the
starting time is 1:20 pm, enter “13.20”.
(3) Press and hold the ZERO button to start logging.
The logging starts.
“MNT1” appears while logging is in progress.
(4) To end the logging, press the BZ.STOP button.
“MNT1” disappears and the logging ends.
•
Once data logging starts, the peak concentration value
every 10 seconds will be logged.
•
The oldest events will be deleted to make room for
the newest when the unit runs out of room for new ones.
•
Data logging automatically stops when the detector is turned
off.
•
Logging date cannot be entered.
•
Refer to the XPS7L data logger software’s instruction
manual (separate document) for the procedure to transfer
log data to a pc.